Assessment of MRI issues at 7 T for 28 implants and other objects.
نویسندگان
چکیده
OBJECTIVE Metallic implants are currently a contraindication for volunteer subjects and patients referred for 7-T examinations because of concerns related to magnetic field interactions and MRI-related heating. Artifacts may also be problematic. Therefore, the purpose of this investigation was to evaluate these MRI issues for 28 implants and other objects in association with a 7-T MR system. MATERIALS AND METHODS Tests were performed at 7 T using standardized procedures to evaluate magnetic field interactions (translational attraction and torque) for all 28 items. MRI-related heating and artifacts were assessed using spin-echo and gradient-echo pulse sequences, respectively, for two aneurysm clips located within a transmit-receive head radiofrequency coil. RESULTS Eight of the 28 items showed magnetic field interactions at levels that could pose risks to human subjects. The two aneurysm clips exhibited heating, but the temperature rise did not exceed 1°C. Artifacts were dependent on the material and dimensions of each aneurysm clip. CONCLUSION These findings show that certain implants and objects may be acceptable for human subjects undergoing MRI examinations at 7 T, whereas others may involve possible risks. This information has important implications for individuals referred for MRI examinations at 7 T.
منابع مشابه
I-45: Advance MRI Sequences in Pelvic Endometriosis
Background: To assess MRI in diagnosing endometriotic lesions, emphasizing T2*weighted imaging efficacy. Materials and Methods: This prospective study of 48 females (22-38 years, average 29.6) clinically suspected of endometriosis from September 2009 to April 2012. MRI was performed with a 1.5 T imager (Siemens) with a body array coil. T1, T2 and T2* weighted (2D-FLASH) sequences were obtained ...
متن کاملMR safety assessment of potential RF heating from cranial fixation plates at 7 T.
PURPOSE The increasing number of clinically oriented MRI studies at 7 T motivates the safety assessment of implants, since many 7 T research sites conservatively exclude all subjects with metallic implants, regardless of type or location. The purpose of this study was to investigate potential RF-induced heating during a 7 T MRI scan using a self-built transmit/receive RF coil in patients with i...
متن کاملApplication of Shape Analysis on 3D Images - MRI of Renal Tumors
The image recognotion and the classification of objects according to the images are more in focus of interests, especially in medicine. A mathematical procedure allows us, not only to evaluate the amount of data per se, but also ensures that each image is pro- cessed similarly. Here in this study, we propose the power of shape analysis, in conjunction with neural networks for reducing white n...
متن کاملAn Investigation of the Dose Distribution from LDR Ir-192 Wires in the Triangular Implants of the Paris System using Polymer Gel Dosimetry
Introduction: Polymer gels are modern dosimeters providing three dimensional dose distributions. These dosimeters can be used in brachytherapy in which the tumor dimension is relatively small and the dose gradient is high. In this study, the ability of the MAGICA polymer gel was investigated for assessing the absolute dose values as well as the dose distribution of low dose rate (LDR) Ir-192 wi...
متن کاملChapter 16 MRI Issues for Implants and Devices
An important aspect of protecting patients from magnetic resonance imaging (MRI)related accidents and injuries involves an understanding of the risks associated with the implants, devices, and other objects (e.g., metallic foreign bodies) that may cause problems in this setting. This requires constant attention and diligence to obtain information and documentation about these items as part of t...