Impact of MLC leaf width on volumetric‐modulated arc therapy planning for head and neck cancers

نویسندگان

  • Caroline Lafond
  • Enrique Chajon
  • Anne Devillers
  • Guillaume Louvel
  • Sandra Toublanc
  • Mickael Olivier
  • Antoine Simon
  • Renaud de Crevoisier
  • Jean‐Pierre Manens
چکیده

This dosimetric study investigated the impact of multileaf collimators (MLC) leaf width in volumetric-modulated arc therapy (VMAT) for head and neck cancers (HNC), either with a "standard" simultaneously integrated boost technique (S-SIB) or with a "dose painting" SIB technique (DP-SIB). HNC patients were planned either with an S-SIB comprising three dose levels, from 56 to 70 Gy (16 patients), or with a DP-SIB comprising five dose levels, from 56 to 84 Gy (8 patients), in 35 fractions. Two VMAT plans were calculated for each SIB technique using two Elekta MLCs: MLCi2 with 10 mm leaf width and Beam Modulator (BM) with 4 mm leaf width. Dose distributions were evaluated by comparing doses on PTVs, main OARs, and healthy tissue, and by comparing conformation indexes. Treatment efficiencies were evaluated by comparing the number of monitor units and the number of needed arcs. Comparisons of the two MLCs depending on the two SIB techniques showed: i) Regarding PTVs: Dmean and D2% on lower doses PTV decreased respectively by 0.5 Gy (p = 0.01) and 0.9 Gy (p = 0.01) with BM than with MLCi2 for S-SIB; no significant difference was found for DP-SIB;ii) Regarding OARs: for spinal cord and brainstem, D2% decreased respectively by 1.2 Gy (p = 0.03) and 4.2 Gy (p = 0.04) with BM than with MLCi2 for S-SIB; for controlateral parotid, D50% decreased by 1.5 Gy (p = 0.01) with BM than with MLCi2 for S-SIB; iii) Regarding treatment efficiency: the number of monitor units was 44% (p = 0.00) and 51% (p = 0.01) higher with BM for S-SIB and DP-SIB, respectively. Two arcs were more frequently needed with BM to reach an acceptable dose distribution. This study demonstrated that Beam Modulator (4 mm leaf width) and MLCi2 (10 mm leaf width) MLCs from Elekta provided satisfactory dose distributions for treatment delivery with VMAT technique for complex HNC cases with standard and dose painting prescriptions. OAR sparing was better with BM, mainly for brainstem and spinal cord. However, delivery efficiency of VMAT plans was better with MLCi2.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Implementation of volumetric arc therapy for head and neck cancer

In volumetric modulated arc therapy (VMAT), treatment is delivered using a continuous arc motion of the gantry with simultaneous variation of mlc position, gantry speed and dose rate. Several manufacturers offer VMAT facilities based on conventional accelerator gantries while others offer tomographic therapy based on ring gantries. A number of centres have begun clinical VMAT treatments in rece...

متن کامل

Beam modeling and VMAT performance with the Agility 160‐leaf multileaf collimator

The Agility multileaf collimator (Elekta AB, Stockholm, Sweden) has 160 leaves of projected width 0.5 cm at the isocenter, with maximum leaf speed 3.5 cms-1. These characteristics promise to facilitate fast and accurate delivery of radiotherapy, particularly volumetric-modulated arc therapy (VMAT). The aim of this study is therefore to create a beam model for the Pinnacle3 treatment planning sy...

متن کامل

Dosimetric impact of multileaf collimator leaf width according to sophisticated grade of technique in the IMRT and VMAT planning for pituitary adenoma lesion

We analyzed the difference in the dosimetric effect between 5-mm and 2.5-mm multileaf collimator (MLC) leaf width according to the sophisticated grades of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). Nineteen patients with pituitary adenomas were selected for this study. The treatment plans were performed according to the size of the MLC (5-mm and 2.5-mm ...

متن کامل

Treatment planning for volumetric modulated arc therapy.

PURPOSE Volumetric modulated arc therapy (VMAT) is a specific type of intensity-modulated radiation therapy (IMRT) in which the gantry speed, multileaf collimator (MLC) leaf position, and dose rate vary continuously during delivery. A treatment planning system for VMAT is presented. METHODS Arc control points are created uniformly throughout one or more arcs. An iterative least-squares algori...

متن کامل

The effect of multileaf collimator leaf width on the radiosurgery planning for spine lesion treatment in terms of the modulated techniques and target complexity

PURPOSE We aim to evaluate the effects of multileaf collimator (MLC) leaf width (5 mm vs. 2.5 mm) on the radiosurgery planning for the treatment of spine lesions according to the modulated techniques (intensity-modulated radiotherapy [IMRT] vs. volumetric-modulated arc therapy [VMAT]) and the complexity of the target shape. METHODS For this study, artificial spinal lesions were contoured and ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2013