Insertion torque versus mechanical resistance of mini-implants inserted in different cortical thickness

نویسندگان

  • Renata de Faria Santos
  • Antonio Carlos de Oliveira Ruellas
  • Daniel Jogaib Fernandes
  • Carlos Nelson Elias
چکیده

OBJECTIVE This study aimed to measure insertion torque, tip mechanical resistance to fracture and transmucosal neck of mini-implants (MI) (Conexão Sistemas de PróteseT), as well as to analyze surface morphology. METHODS Mechanical tests were carried out to measure the insertion torque of MIs in different cortical thicknesses, and tip mechanical resistance to fracture as well as transmucosal neck of MIs. Surface morphology was assessed by scanning electron microscopy (SEM) before and after the mechanical tests. RESULTS Values of mechanical resistance to fracture (22.14 N.cm and 54.95 N.cm) were higher and statistically different (P < 0.05) from values of insertion torque for 1-mm (7.60 N.cm) and 2-mm (13.27 N.cm) cortical thicknesses. Insertion torque was statistically similar (P > 0.05) to torsional fracture in the tip of MI (22.14 N.cm) when 3 mm cortical thickness (16.11 N.cm) and dense bone (23.95 N.cm) were used. Torsional fracture of the transmucosal neck (54.95 N.cm) was higher and statistically different (P < 0.05) from insertion torsional strength in all tested situations. SEM analysis showed that the MIs had the same smooth surface when received from the manufacturer and after the mechanical tests were performed. Additionally, no significant marks resulting from the manufacturing process were observed. CONCLUSION All mini-implants tested presented adequate surface morphology. The resistance of mini-implants to fracture safely allows placement in 1 and 2-mm cortical thickness. However, in 3-mm cortical thickness and dense bones, pre-drilling with a bur is recommended before insertion.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Insertion torque, pull-out strength and cortical bone thickness in contact with orthodontic mini-implants at different insertion angles.

This study aimed to evaluate biomechanical behaviour of inclined orthodontic mini-implants by analyzing its insertion torque (IT), axial pull-out strength (APS), and cortical bone thickness in contact with mini-implant (CBTC). A total of 102 mini-implants were inserted at 90 degree, 60 degree, and 45 degree to the surface of synthetic bone. Peak IT was measured, and the mini-implants were align...

متن کامل

Influence of cortical thickness on the stability of mini-implants with microthreads.

The objective of this study was to assess the influence of cortical thickness and bone density on the insertion torque of a mini-implant (MI) with microthreads. Mini-implants with lengths of 6 and 8 mm in the active part were inserted into synthetic bone blocks (polyurethane resin). The density of these blocks was 20 pounds per cubic foot (pcf), simulating bone marrow, and that of blocks 1, 2, ...

متن کامل

Influence of Screw Length and Bone Thickness on the Stability of Temporary Implants

The purpose of this work was to study the influence of screw length and bone thickness on the stability of temporary implants. A total of 96 self-drilling temporary screws with two different lengths were inserted into polyurethane blocks (n = 66), bovine femurs (n = 18) and rabbit tibia (n = 12) with different cortical thicknesses (1 to 8 mm). Screws insertion in polyurethane blocks was assiste...

متن کامل

Geometrical design characteristics of orthodontic mini-implants predicting maximum insertion torque

OBJECTIVE To determine the unique contribution of geometrical design characteristics of orthodontic mini-implants on maximum insertion torque while controlling for the influence of cortical bone thickness. METHODS Total number of 100 cylindrical orthodontic mini-implants was used. Geometrical design characteristics of ten specimens of ten types of cylindrical self-drilling orthodontic mini-im...

متن کامل

Insertion Torques of Self-Drilling Mini-Implants in Simulated Mandibular Bone: Assessment of Potential for Implant Fracture.

PURPOSE Fracture of orthodontic mini-implants during insertion is a limiting factor for their clinical success. The purpose of this study was to determine the fracture potential of commonly used self-drilling orthodontic mini-implants when placed into simulated thick, dense mandibular bone. MATERIALS AND METHODS Six mini-implant systems were assessed for the potential for fracture (Aarhus, Me...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 19  شماره 

صفحات  -

تاریخ انتشار 2014