Experimental and Monte Carlo evaluation of an ionization chamber in a 60 Co beam
نویسندگان
چکیده
Recently a special parallel-plate ionization chamber was developed and characterized at the Instituto de Pesquisas Energéticas e Nucleares. The operational tests presented results within the recommended limits. In order to determine the influence of some components of the ionization chamber on its response, Monte Carlo simulations were carried out. The experimental and simulation results pointed out that the dosimeter evaluated in the present work has favorable properties to be applied to 60 Co dosimetry at calibration laboratories.
منابع مشابه
An Efficiency Studying of an Ion Chamber Simulation Using Vriance Reduction Techniques with EGSnrc
Background: Radiotherapy is an important technique of cancer treatment using ionizing radiation. The determination of total dose in reference conditions is an important contribution to uncertainty that could achieve 2%. The source of this uncertainty comes from cavity theory that relates the in-air cavity dose and the dose to water. These correction factors are determined from Monte Carlo calcu...
متن کاملSU-E-T-15: Small and Nonstandard Photon Field Dosimetry Characterization Using Monte Carlo Methods.
PURPOSE To quantify uncertainty reduction in small photon field dosimetry through characterization of ionization chambers and calibration conditions using detailed Monte Carlo methods benchmarked against NIST-traceable measurements. METHODS Phase space profiles were obtained using detailed EGSnrc Monte Carlo models for a Varian 6 MV photon linear accelerator, and a NIST-traceable cobalt-60 te...
متن کاملInvestigation of the air kerma response of spherical ionization chambers for unfolded pulse height distributions of (60)Co and (137)Cs using the EGS4 Monte Carlo code.
Data required for the determination of the absolute air kerma rate for (60)Co and (137)Cs gamma-rays using spherical cavity chambers were calculated using the EGS4 Monte Carlo system. Mass energy-absorption coefficient ratio and the stopping power ratio were calculated for a 10 cm(3) primary standard graphite-walled ionization chamber from the unfolded energy pulse height distributions of (60)C...
متن کاملSmall photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملMonte Carlo estimation of electron contamination in a 18 MV clinical photon beam
Background: The electron contamination may reduce or even diminish the skin sparing property of the megavoltage beam. The detailed characteristics of contaminant electrons are presented for different field sizes and cases. Materials and Methods: The Monte Carlo code, MCNPX, has been used to simulate 18 MV photon beam from a Varian Linac-2300 accelerator. All dose measurements were car...