Sensitive high-pressure liquid chromatographic assay for amphotericin B which incorporates an internal standard.
نویسندگان
چکیده
We developed a rapid, sensitive, high-pressure liquid chromatographic (HPLC) procedure which incorporates a commercially available internal standard, 1-amino-4-nitronaphthalene, to measure amphotericin B in serum. Recovery was quantitative (greater than or equal to 90%), and the standard curve was linear from 0.04 to at least 10.0 micrograms/ml. The reproducibility of the assay was good, with intrarun coefficients of variation from 2.0 to 6.8% and interrun coefficients of variation from 4.9 to 10.0%. Comparison by linear regression analysis of the HPLC assay with an agar well diffusion bioassay gave a correlation coefficient of 0.942, with the HPLC assay exhibiting greater precision and sensitivity. No interference was encountered from over 20 drugs and three amphotericin B analogs. However, serum specimens that contained high concentrations of conjugated bilirubin (greater than 3 mg/dl) produced interfering peaks in both this assay and other previously reported HPLC assays for amphotericin B. We also describe a solid-phase extraction procedure which effectively removes this interference and uses an alternative internal standard (N-acetyl amphotericin B).
منابع مشابه
Simple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma
A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-...
متن کاملSimple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma
A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-...
متن کاملRapid High Performance Liquid Chromatographic Determination of Risperidone in Human Plasma
A simple, rapid and sensitive high-performance liquid chromatographic (HPLC) method for the determination of risperidone in human plasma was developed. An HPLC system based on a Nucleosil C8 column (150?4 mm) and a UV detector (?= 280 nm) were used. A mixture of sodium dihydrogen phosphate buffer-acetonitrile (55:45, v/v) adjusted to pH 6.0 at a flow rate of 1.5 ml min-1 was used as mobile phas...
متن کاملRapid High Performance Liquid Chromatographic Determination of Risperidone in Human Plasma
A simple, rapid and sensitive high-performance liquid chromatographic (HPLC) method for the determination of risperidone in human plasma was developed. An HPLC system based on a Nucleosil C8 column (150?4 mm) and a UV detector (?= 280 nm) were used. A mixture of sodium dihydrogen phosphate buffer-acetonitrile (55:45, v/v) adjusted to pH 6.0 at a flow rate of 1.5 ml min-1 was used as mobile phas...
متن کاملA High-Performance Liquid Chromatographic Assay for the Determination of Losartan in Plasma
A rapid and sensitive HPLC method was developed for determination of losartan in plasma. Losartan was extracted from plasma by a two-step extraction procedure using chloroform as extracting solvent in acidic medium. HPLC analysis was performed on a cyano reversed-phase column using phosphate buffer (pH 4.3), acetonitrile (750:250, v/v) as mobile phase with a flow rate of 0.9 mL/min. Sodium dicl...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Antimicrobial agents and chemotherapy
دوره 29 4 شماره
صفحات -
تاریخ انتشار 1986