BACKGROUND There is minimal experience with less rigid syndesmotic fixation devices which may approximate the normal distal tibio-fibular mechanics during healing. This study evaluates the ability of a FiberWire-button implant (Arthrex, Naples, FL) to maintain syndesmotic reduction as compared with a metallic screw. METHODS Ten matched fresh-frozen cadaveric ankle pairs with intact ligaments were tested (12.5 Nm external rotation force) to establish physiologic syndesmotic diastasis. The same force was applied to the ankles after sectioning of the syndesmotic and deltoid ligaments. Within the pairs, each limb was randomized to receive a FiberWire-button implant or a metallic screw (Synthes, Paoli, PA); the ankles were tested for syndesmotic diastasis with progressive external rotation force, from 2.5 Nm to 25 Nm (or failure). RESULTS There was no significant difference in diastasis amongst pairs with intact or sectioned syndesmosis (p=0.64 and p=0.80, respectively). There was a significantly greater diastasis in the FiberWire-button group at all external rotation loads (p<0.0001). Nine of the ten pairs failed (all through fracture of the distal fibula). There were no hardware failures. The metallic screw group failed at a lower load (mean 15 Nm) compared to the FiberWire-button group (mean 18 Nm, p=0.0004). The metallic screw group maintained syndesmotic reduction up to 5 Nm of force. CONCLUSIONS The FiberWire-button was unable to maintain syndesmotic reduction of the ankles at any of the forces applied. The ankles fixed with the FiberWire-button demonstrated significantly greater widening of the syndesmosis compared to the screw, at all loads. CLINICAL RELEVANCE The FiberWire-button implant may not maintain adequate ankle syndesmotic reduction in the immediate post-operative period relative to a metallic screw.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical Comparison of Distal Tibiofibular Syndesmotic Joint Fixation: Novel Suture Construct versus Metallic Screw Construct in Syndesmotic Injury Model

INTRODUCTION: Syndesmotic injuries of the ankle commonly occur via an external rotation force applied to the ankle joint. The effects of rigid or non-rigid fixation of the syndesmotic injury can be assessed by evaluating the three-dimensional kinematics behavior of tibiofibular diastasis. The specific aims of this study were: 1) to determine the physiological range of motion of the distal tibio...

متن کامل

A Novel Suture Button Construct for Acute Ankle Syndesmotic Injuries; A Prospective Clinical and Radiological Analysis.

  Background: The importance of the syndesmosis in ankle stability is well recognized. Numerous means of fixation have been described for syndesmotic injuries including the suture button technique. Significant cost limits the use the commercially available options. We, therefore, designed a cheap and readily available alternative construct. We aim to assess the results of using a novel suture-...

متن کامل

Suture Button Fixation for Syndesmosis Injuries: Review of the Literature

Syndesmosis injuries -also known as high ankle sprainshave an increasing popularity because of an increased awareness of diagnosis. Different from lateral ankle sprains, mechanism of injury mostly involves external rotation, eversion and excessive dorsiflexion. Other mechanisms can also cause sydesmosis injuries. Because of high levels of missed injuries the real prevalence of syndesmosis injur...

متن کامل

A systematic review of suture-button versus syndesmotic screw in the treatment of distal tibiofibular syndesmosis injury

BACKGROUND Now, using a suture-button device to treat distal tibiofibular syndesmotic injuries is overwhelming due to its advantages over screw fixation. Current systematic review was conducted to make a comparison between suture-button fixation and traditionally screw fixation in the treatment of syndesmotic injuries. The outcomes included functional outcomes, implant removal, implant failure,...

متن کامل

Medial malleolar fractures: a biomechanical study of fixation techniques.

Fracture fixation of the medial malleolus in rotationally unstable ankle fractures typically results in healing with current fixation methods. However, when failure occurs, pullout of the screws from tension, compression, and rotational forces is predictable. We sought to biomechanically test a relatively new technique of bicortical screw fixation for medial malleoli fractures. Also, the AO gro...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Foot & ankle international

دوره 29 4  شماره 

صفحات  -

تاریخ انتشار 2008