New method for obtaining position and time structure of source in HDR remote afterloading brachytherapy unit utilizing light emission from scintillator
نویسندگان
چکیده
When using a HDR remote afterloading brachytherapy unit, results of treatment can be greatly influenced by both source position and treatment time. The purpose of this study is to obtain information on the source of the HDR remote afterloading unit, such as its position and time structure, with the use of a simple system consisting of a plastic scintillator block and a charge-coupled device (CCD) camera. The CCD camera was used for recording images of scintillation luminescence at a fixed rate of 30 frames per second in real time. The source position and time structure were obtained by analyzing the recorded images. For a preset source-step-interval of 5 mm, the measured value of the source position was 5.0 +/- 1.0 mm with a pixel resolution of 0.07 mm in the recorded images. For a preset transit time of 30 s, the measured value was 30.0 +/- 0.6 s, when the time resolution of the CCD camera was 1/30 s. This system enables us to obtain the source dwell time and movement time. Therefore, parameters such as 192Ir source position, transit time, dwell time, and movement time at each dwell position can be determined quantitatively using this plastic scintillator-CCD camera system.
منابع مشابه
Independent assessment of source transit time for the BEBIG SagiNova® high dose rate brachytherapy afterloader
Introduction: The dwell time and transit time components contribute to the overall delivered dose to patients in high dose rate (HDR) brachytherapy treatments. The transit time results from source entry and exit as well as source movements between dwell positions. It depends on various parameters such as the source speed profile, source indexer distance to dwell position, and ...
متن کاملEvaluation of Dose Distribution Accuracy in HDR Brachytherapy of Esophagus Cancer Based on MRI Normoxic Polymer Gel Dosimetry
Introduction: The purpose of this work was to study the ability of MRI normoxic polymer gel dosimetry for evaluating the dose distribution in HDR brachytherapy of esophagial cancer at Imam Reza brachytherapy center (Mashhad, Iran). Materials and Methods: Initially, 2liters of normoxic gel (MAGIC) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phanto...
متن کاملRectal and Bladder Dose Measurements in the Intracavitary Applications of Cervical Cancer Treatment with HDR Afterloading System: Comparison of TPS Data with MOSFET Detector
Background: Intracavitary brachytherapy plays a major role in management of cervical carcinoma. Assessment of dose received by OAR’s therefore becomes crucial for the estimation of radiation toxicities in HDR brachytherapy. Objective: Purpose of this study is to evaluate the role of in vivo dosimetry in HDR brachytherapy and to compare actual doses delivered to OAR’ s with those calculate...
متن کاملUsing mean dose rate to compare relative dosimetric efficiency with respect to source type and source change schedules for HDR brachytherapy°
Remote afterloading devices used for high-dose-rate (HDR) brachytherapy may be supplied with different sources, and these sources typically have differing initial source strengths. In addition, the proposed frequency for source changes may also vary, depending upon the source type. Dosimetric parameters unique to each source are often used to compare source types. However, when considering the ...
متن کاملDose error from deviation of dwell time and source position for high dose-rate 192Ir in remote afterloading system
The influence of deviations in dwell times and source positions for (192)Ir HDR-RALS was investigated. The potential dose errors for various kinds of brachytherapy procedures were evaluated. The deviations of dwell time ΔT of a (192)Ir HDR source for the various dwell times were measured with a well-type ionization chamber. The deviations of source position ΔP were measured with two methods. On...