Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

نویسندگان

  • Stephen J Vincent
  • Roslyn A Vincent
  • David Shields
  • Graham A Lee
چکیده

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients including only eyes under medical treatment for glaucoma. METHODS Measurement by Goldmann applanation tonometry, iCare rebound tonometry and 7CR non-contact tonometry. MAIN OUTCOME MEASURES Intraocular pressure. RESULTS There were strong correlations between the intraocular pressure measurements obtained with Goldmann and both the rebound and non-contact tonometers (Spearman r-values ≥ 0.79, P < 0.001). However, there were small, statistically significant differences between the average readings for each tonometer. For the rebound tonometer, the mean intraocular pressure was slightly higher compared with the Goldmann applanation tonometer in the right eyes (P = 0.02), and similar in the left eyes (P = 0.93); however, these differences did not reach statistical significance. The Goldmann correlated measurements from the non-contact tonometer were lower than the average Goldmann reading for both right (P < 0.001) and left (P > 0.01) eyes. The corneal compensated measurements from the non-contact tonometer were significantly higher compared with the other tonometers (P ≤ 0.001). CONCLUSIONS The iCare rebound tonometer and the 7CR non-contact tonometer measure intraocular pressure in fundamentally different ways to the Goldmann applanation tonometer. The resulting intraocular pressure values vary between the instruments and will need to be considered when comparing clinical versus home acquired measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Correlation between diurnal variation of intraocular pressure, ocular pulse amplitude and corneal structural properties.

PURPOSE The purpose of this study was to evaluate the fluctuation of intraocular pressure measurements obtained by Goldmann applanation tonometry, dynamic contour tonometry, and corneal compensated non-contact tonometry during office hours in glaucoma and healthy participants. This study also aims at correlating the intraocular pressure fluctuations with fluctuations of corneal hysteresis, cent...

متن کامل

Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients

This study aims to determine the agreement of patient-measured intraocular pressure (IOP) using rebound tonometry with ophthalmologist-measured IOP using Goldmann applanation tonometry (GAT). Fifty-three glaucoma patients used rebound tonometry (Icare ONE, Icare Finland Oy., Finland) to measure their own IOP in ambient environments for 1 week, 5 times per day. Clinic IOP measurements were perfo...

متن کامل

Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

PURPOSE (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer. METHODS A...

متن کامل

Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study

OBJECTIVE To test agreement of two methods to measure intraocular pressure (IOP): rebound tonometry (RBT) and gold standard Goldmann applanation tonometry (GAT) in children with glaucoma. DESIGN Observational prospective cohort study. SETTING Tertiary paediatric glaucoma clinic at a single centre. PARTICIPANTS 102 individuals attending a paediatric glaucoma clinic, mean (SD) age 11.85 (3....

متن کامل

A comparison between rebound and Goldmann's tonometers in screening of patients for glaucoma.

Goldmann applanation tonometer (GAT) has been the gold standard for intraocular pressure (IOP) measurement for over five decades. However, the last few years have seen lots of research in an effort to develop the ultimate tonometer. One of the outcomes of that effort is the rebound tonometer (RBT) (Icare; Tiolat Oy, Helsinki, Finland), which has appeared in clinical practice after extensive res...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical & experimental ophthalmology

دوره 40 4  شماره 

صفحات  -

تاریخ انتشار 2012