Comparison of the Colvard, Procyon, and Neuroptics pupillometers for measuring pupil diameter under low ambient illumination.

نویسندگان

  • Maurice Schallenberg
  • Valerie Bangre
  • Klaus-Peter Steuhl
  • Stephan Kremmer
  • J Michael Selbach
چکیده

PURPOSE To compare three different pupillometers (Colvard, Procyon, and Neuroptics) for determining pupil diameter at 0.04 and 0.4 lux ambient illumination. METHODS In 92 eyes of 46 healthy volunteers, pupil diameter was measured at 0.04 and 0.4 lux. After dark adaptation for 2 minutes, measurements were performed with each device by two examiners. Interobserver agreement, instrument agreement, and repeatability were analyzed. RESULTS Mean pupil diameter was 6.63+/-0.68 mm, 6.24+/-1.01 mm, and 6.99+/-0.67 mm at 0.04 lux and 6.22+/-0.74, 4.64+/-1.04, and 6.73+/-0.72 mm at 0.4 lux with the Colvard, Procyon, and Neuroptics pupillometers, respectively. The interobserver disagreement ranged within narrower limits for the Colvard (0.04 lux: -1.0 to 0.5 mm; 0.4 lux: -0.75 to 1.0 mm) and Neuroptics (0.04 lux: -1.0 to 0.5 mm; 0.4 lux: -1.7 to 0.7 mm) than for the Procyon (0.04 lux: -0.74 to 1.14 mm; 0.4 lux -1.82 to 2.4 mm) under both light conditions. Instrument agreement ranged within narrower limits for the Colvard versus Neuroptics (0.04 lux: -1.3 to 0.75 mm; 0.4 lux: -1.55 to 1.40 mm) than for the Neuroptics versus Procyon (0.04 lux: -1.06 to 2.69 mm; 0.4 lux: 0.18 to 3.69 mm) or Colvard versus Procyon (0.04 lux: -0.63 to 2.60 mm; 0.4 lux: -0.32 to 3.13 mm) at both light levels. At 0.04 lux, repeatability showed no measurement difference outside +/-0.5 mm for the Colvard and Neuroptics; for the Procyon, 25% of consecutive measurements showed a difference >+/-0.5 mm. At 0.4 lux, 2.5% of consecutive measurements for the Colvard and 5% for the Neuroptics differed by >+/-0.5 mm; for the Procyon, 13% of measurements differed by more than this amount. CONCLUSIONS Pupil diameters under both light conditions were largest with the Neuroptics pupillometer and smallest with the Procyon. The most "examiner independent" Procyon pupillometer performed poorly. The underestimation of the pupil diameter might have severe consequences for refractive surgery patients. The Neuroptics pupillometer showed a high interobserver agreement and repeatability and therefore high safety.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of measurement of pupil sizes among the colvard pupillometer, procyon pupillometer, and NIDEK OPD-scan.

PURPOSE To compare pupil sizes measured with the Colvard pupillometer, Procyon pupillometer, and NIDEK OPD-Scan. METHODS Pupil diameter was measured in 90 consecutive eyes from 55 patients under mesopic and scotopic light conditions with all three instruments. RESULTS The mean scotopic pupil diameter was 6.3+/-0.98 mm with the Colvard pupillometer and 6.45+/-1.01 mm with the Procyon pupillo...

متن کامل

Evaluation of pupil dynamics after implantation of artisan phakic intraocular lenses.

PURPOSE To compare pre- and postoperative horizontal and vertical pupil diameters after horizontal Artisan phakic intraocular lens (PIOL) (Ophtec BV, Groningen, The Netherlands) implantation for correction of myopia. METHODS The pre- and postoperative scotopic pupil diameters measured by the Colvard pupillometer (Oasis Medical, Glendora, Calif) were compared in 71 eyes after Artisan PIOL impl...

متن کامل

Suitability of open-field autorefractors as pupillometers and instrument design effects.

AIM To determine the agreement and repeatability of the pupil measurement obtained with VIP-200 (Neuroptics), PowerRef II (Plusoptix), WAM-5500 (Grand Seiko) and study the effects of instrument design on pupillometry. METHODS Forty patients were measured twice in low, mid and high mesopic. Repeatability was analyzed with the within-subject standard deviation (Sw) and paired t-tests. Agreement...

متن کامل

Pre-excimer laser and post-excimer laser refractive surgery measurements of scotopic pupil diameter using 2 pupillometers.

PURPOSE To compare a digital infrared pupillometer with a handheld light amplification pupillometer for measuring scotopic pupil size and to evaluate if the postoperative refractive changes of the cornea can influence pupil measurements. DESIGN Prospective noncomparative interventional case series. PARTICIPANTS One hundred eyes, 50 myopic (mean spherical equivalent [SE] refraction [+/- stan...

متن کامل

Colvard pupillometer measurement of scotopic pupil diameter in emmetropes and myopes.

PURPOSE To prospectively compare the scotopic pupil size between emmetropes and myopes using a Colvard pupillometer. METHODS The pupil diameters of 55 normal subjects and 55 healthy myopic subjects were measured with the Colvard pupillometer in a low-light situation that simulated the level of light encountered while driving at night. RESULTS The mean (+/- SD) age of the emmetropic subjects...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of refractive surgery

دوره 26 2  شماره 

صفحات  -

تاریخ انتشار 2010