Structural integrity of balloon-expandable stents after transcatheter aortic valve replacement: assessment by multidetector computed tomography.
نویسندگان
چکیده
OBJECTIVES This study sought to evaluate the structural integrity of balloon-expandable stents used in transcatheter aortic valve replacement. BACKGROUND Underexpansion, deformation, or fracture of stent frames may affect transcatheter heart valve (THV) function and durability. METHODS Patients >1 year after transcatheter aortic valve replacement underwent multidetector computed tomography. Geometry of the stent frame was assessed for circularity; eccentricity; minimum and maximum external diameter; and expansion at the inflow, mid-stent, and outflow levels, as well as for stent fracture. THV noncircularity was defined as stent eccentricity >10% (1 - minimum diameter/maximum diameter) and THV underexpansion when expansion <90% (multidetector computed tomography derived external valve area/nominal external valve area). Echocardiography was performed after implantation and annually. RESULTS Fifty patients underwent multidetector computed tomography at an average of 2.5 ± 0.9 years after transcatheter aortic valve replacement (35 Sapien, 8 Sapien XT, and 7 Cribier-Edwards valves [all Edwards Lifesciences, Irvine, California). The mean external diameter for the 23- and 26-mm valves was 23.3 ± 0.9 mm and 25.9 ± 0.9 mm, respectively. Circularity was present in 96% (48 of 50) and median eccentricity was 2.0% (interquartile range: 1.2% to 3.0%). Mean THV expansion was 104.1 ± 7.4%, which increased from stent inflow to outflow (100.8 ± 7.6% vs. 108.1 ± 6.9%, p < 0.001). Stent fracture was not observed. Underexpanded valves (8% [4 of 50]) and noncircular valves (4% [2 of 50]) demonstrated stable hemodynamic function on annual echocardiography. CONCLUSIONS Balloon-expandable aortic valves have excellent rates of circularity with low eccentricity and maintain full expansion without stent fracture at an average 2.5 years after implantation.
منابع مشابه
Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement.
BACKGROUND Aortic root rupture is a major concern with balloon-expandable transcatheter aortic valve replacement (TAVR). We sought to identify predictors of aortic root rupture during balloon-expandable TAVR by using multidetector computed tomography. METHODS AND RESULTS Thirty-one consecutive patients who experienced left ventricular outflow tract (LVOT)/annular/aortic contained/noncontained...
متن کاملTranscatheter aortic valve replacement with the SAPIEN 3: a new balloon-expandable transcatheter heart valve.
OBJECTIVES The aim of this study was to demonstrate the first-in-human feasibility and short-term clinical outcomes with a new balloon-expandable transcatheter heart valve (THV). BACKGROUND The SAPIEN 3 (S3) THV incorporates a paravalvular sealing system, an active 3-dimensional coaxial positioning catheter, and is compatible with a 14-F expandable sheath. METHODS The S3 THV was implanted i...
متن کاملProsthesis oversizing in balloon-expandable transcatheter aortic valve implantation is associated with contained rupture of the aortic root.
BACKGROUND To retrospectively investigate the potential cause of contained rupture of the aortic root in balloon-expandable transcatheter aortic valve implantation (TAVI) by means of pre- and postinterventional multislice computed tomography. METHODS AND RESULTS Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm(2)) underwent balloon-expandable TAVI using the Edwar...
متن کاملIs time of renal hypoperfusion an important variable in determining response to renal artery revascularization?
3. Gurvitch R, Webb JG, Yuan R, et al. Aortic annulus diameter determination by multidetector computed tomography: reproducibility, applicability, and implications for transcatheter aortic valve implantation. J Am Coll Cardiol Intv 2011;4:1235–45. 4. Achenbach S, Delgado V, Hausleiter J, Schoenhagen P, Min JK, Leipsic JA. SCCT expert consensus document on computed tomography imaging before tran...
متن کاملQuantity and location of aortic valve complex calcification predicts severity and location of paravalvular regurgitation and frequency of post-dilation after balloon-expandable transcatheter aortic valve replacement.
OBJECTIVES This study sought to determine the impact of quantity and location of aortic valve calcification (AVC) on paravalvular regurgitation (PVR) and rates of post-dilation (PD) immediately after transcatheter aortic valve replacement (TAVR). BACKGROUND The impact of AVC in different locations within the aortic valve complex is incompletely understood. METHODS This study analyzed 150 pa...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- JACC. Cardiovascular interventions
دوره 5 5 شماره
صفحات -
تاریخ انتشار 2012