Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis
نویسندگان
چکیده
OBJECTIVES The present study used strain gauge analysis to perform an in vitro evaluation of the effect of axial loading on 3 elements of implant-supported partial fixed prostheses, varying the type of prosthetic cylinder and the loading points. MATERIAL AND METHODS Three internal hexagon implants were linearly embedded in a polyurethane block. Microunit abutments were connected to the implants applying a torque of 20 Ncm, and prefabricated Co-Cr cylinders and plastic prosthetic cylinders were screwed onto the abutments, which received standard patterns cast in Co-Cr alloy (n=5). Four strain gauges (SG) were bonded onto the surface of the block tangentially to the implants, SG 01 mesially to implant 1, SG 02 and SG 03 mesially and distally to implant 2, respectively, and SG 04 distally to implant 3. Each metallic structure was screwed onto the abutments with a 10 Ncm torque and an axial load of 30 kg was applied at five predetermined points (A, B, C, D, E). The data obtained from the strain gauge analyses were analyzed statistically by RM ANOVA and Tukey's test, with a level of significance of p<0.05. RESULTS There was a significant difference for the loading point (p=0.0001), with point B generating the smallest microdeformation (239.49 µε) and point D the highest (442.77 µε). No significant difference was found for the cylinder type (p=0.748). CONCLUSIONS It was concluded that the type of cylinder did not affect in the magnitude of microdeformation, but the axial loading location influenced this magnitude.
منابع مشابه
Effect of Three Types of Temporary Luting Cements and Abutment Surface Sandblasting on Retentive Strength of Implant-Supported Fixed Prostheses
Background and Aim: Achieving appropriate retention and easy retrieva bility at the same time is challenging in implant-supported fixed par tial dentures (FPDs). Researchers have always been in search of a temporary cement to improve the retentive strength. The aim of this in vitro study was to evaluate the effect of three different temporary ce ments and sandblasting of the abutment surface on...
متن کاملCase Report: Implant-Supported Removable Partial Denture In A Mandibular Kennedy CL II: A Clinical Report
Abstract The implant-supported removable partial dentures should be an alternative to conventional removable partial dentures and implant-supported fixed partial prostheses when the implant’s insertion is confined by bone height and thickness. Placing a single implant in the posterior region would modify the Kennedy Class II configuration to a Kennedy Class III and increase the stability a...
متن کاملPeri-implant biomechanical responses to standard, short-wide, and double mini implants replacing missing molar supporting hybrid ceramic or full-metal crowns under axial and off-axial loading: an in vitro study
BACKGROUND The aim of this study was to evaluate the biomechanical response of the peri-implant bone to standard, short-wide, and double mini implants replacing missing molar supporting either hybrid ceramic crowns (Lava Ultimate restorative) or full-metal crowns under two different loading conditions (axial and off-axial loading) using strain gauge analysis. METHODS Three single-molar implan...
متن کاملتعیین توزیع تنش استخوان اطراف ایمپلنتهای فک پایین در طرحهای مختلف پروتز به هنگام Mandibular- Flexure
Statement of Problem: In the treatment of edentulous patients with implant supported fixed partial dentures several factors such as implant numbers, implant position, superstructure pattern and cantilever length must be considered. Mandibular flexture in function exerts forces in peri-implant bone, however this phenomenon has received little attention. Purpose: The goal of this finite element a...
متن کاملمطالعه تنشهای اطراف ایمپلنت به روش اجزای محدود در بریجهای با ساپورت دندان/ایمپلنت در شرایط اتصال سخت و غیرسخت در پروتزهای سمان شونده
Background and Aims: Freestanding fixed partial prosthesis is considered the first choice whenever possible. However, anatomical limitations for implants and other reasons may create situation in which it would be preferable to connect the implants to teeth. A biomechanical dilemma in a tooth/implant-supported system comes from dissimilar mobility. This disparity cause the bridge to function as...
متن کامل