Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus

نویسندگان

  • M J Fokkert
  • P R van Dijk
  • M A Edens
  • S Abbes
  • D de Jong
  • R J Slingerland
  • H J G Bilo
چکیده

OBJECTIVE To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. RESEARCH DESIGN AND METHODS 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2). RESULTS FSL-CGM arm sensor readings showed 85.5% of paired readings falling within Clarke Error Grid (ISO 15197:2013) zone A when compared with StatStrip. For FSL-CGM abdomen and FSLC, these percentages were 64% and 98%, respectively. The overall correlation of FSL-CGM in the arm and the StatStrip indicates a performance with lower results with the FSL-CGM in the arm than expected based on the StatStrip in the lower glucose ranges, and higher results than expected in the higher ranges. Following a standardized glucose load, a slower rise in glucose level was observed for FSL-CGM arm as compared with PCA-HK, StatStrip, FSLC, and iPro2 during the first 45-60 min after glucose load ingestion. CONCLUSIONS Certain matters need attention while using the FSL-CGM in daily life including the observed lower values in the lower ranges, and the underestimation of the effect of a meal on glucose response. These effects of such deviations can partly be overcome by optimizing the available user instructions. TRIAL REGISTRATION NUMBER TC5348; results.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes

BACKGROUND In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system. METHODS Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10-14 days and measured capillary blood glucose levels with t...

متن کامل

Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes

BACKGROUND Accuracy of the FreeStyle Libre™ Flash Glucose Monitoring System has not been evaluated in pregnant women with diabetes. The aim of this study was to determine accuracy (compared to self-monitoring of blood glucose [SMBG]), clinical safety, and acceptability of the FreeStyle Libre System when used at home by this population. MATERIALS AND METHODS Seventy-four participants, with typ...

متن کامل

The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System

INTRODUCTION The purpose of the study was to evaluate the performance and usability of the FreeStyle(®) Libre™ Flash glucose monitoring system (Abbott Diabetes Care, Alameda, CA) for interstitial glucose results compared with capillary blood glucose results. MATERIALS AND METHODS Seventy-two study participants with type 1 or type 2 diabetes were enrolled by four U.S. clinical sites. A sensor ...

متن کامل

The performance of flash glucose monitoring in critically ill patients with diabetes.

OBJECTIVE Frequent glucose monitoring may improve glycaemic control in critically ill patients with diabetes. We aimed to assess the accuracy of a novel subcutaneous flash glucose monitor (FreeStyle Libre [Abbott Diabetes Care]) in these patients. METHODS We applied the FreeStyle Libre sensor to the upper arm of eight patients with diabetes in the intensive care unit and obtained hourly flash...

متن کامل

Flash Glucose Monitoring: The Future Is Here

We cannot believe that it has been more than two decades since we first evaluated real-time continuous glucose monitoring (CGM)—(GlucoWatch ) in patients with type 1 diabetes (T1D). The first real-time CGM (GlucoWatch), which uses a reversed iontophoresis to measure interstitial glucose, was approved to be used for 12 h by the Food and Drug Administration (FDA) in 1999. Since then, several adva...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2017