Accuracy of reconstruction of the hip using computerised three-dimensional pre-operative planning and a cementless modular neck.
نویسندگان
چکیده
Pre-operative computerised three-dimensional planning was carried out in 223 patients undergoing total hip replacement with a cementless acetabular component and a cementless modular-neck femoral stem. Components were chosen which best restored leg length and femoral offset. The post-operative restoration of the anatomy was assessed by CT and compared with the pre-operative plan. The component implanted was the same as that planned in 86% of the hips for the acetabular implant, 94% for the stem, and 93% for the neck-shaft angle. The rotational centre of the hip was restored with a mean accuracy of 0.73 mm (SD 3.5) craniocaudally and 1.2 mm (SD 2) laterally. Limb length was restored with a mean accuracy of 0.3 mm (SD 3.3) and femoral offset with a mean accuracy of 0.8 mm (SD 3.1). This method appears to offer high accuracy in hip reconstruction as the difficulties likely to be encountered when restoring the anatomy can be anticipated and solved pre-operatively by optimising the selection of implants. Modularity of the femoral neck helped to restore the femoral offset and limb length.
منابع مشابه
Prediction Of The Primary Stem Stability For Total Hip Arthroplasty Using Numerical Modal Analysis Based On Preoperative Computerised Three- dimensional Planning
Introduction: Cementless total hip arthroplasty involves a press-fit condition between the stem and the femur. The primary stem stability is crucial in order to achieve bone in-growth and good functional outcomes especially in the early post-operative period. The stem stability is usually assessed intuitively by the surgeon at the time of the surgical procedure with an inherent risk of size err...
متن کاملComparisons of preoperative three-dimensional planning and surgical reconstruction in primary cementless total hip arthroplasty.
Reconstruction of important parameters such as femoral offset and torsion is inaccurate, when templating is based on plain x-rays. We evaluate intraoperative reproducibility of pre-operative CT-based 3D-templating in a consecutive series of 50 patients undergoing primary cementless THA through an anterior approach. Pre-operative planning was compared to a postoperative CT scan by image fusion. ...
متن کاملTotal Hip Arthroplasty with Modular Stem for Dysplastic Hips in South Asian Population
Introduction: Optimum component positioning and orientation is required to optimize the functional result during total hip arthroplasty for dysplastic hips. Patients and methods: Sixty-two patients (66 hips) including 33 males and 29 females underwent total hip arthroplasty using modular stem prosthesis at an average age of 40.6 years (range 17 to 49 years). Nineteen hips were clas...
متن کاملTotal hip arthroplasty with a sliding iliac graft for acetabular dysplasia.
We describe a new technique of reconstruction of the deficient acetabulum in cementless total hip arthroplasty. The outer iliac table just above the deficient acetabulum is osteotomised and slid downwards. We have termed this an iliac sliding graft. Between October 1997 and November 2001, cementless total hip arthroplasty with an iliac sliding graft was performed on 19 patients (19 hips) with a...
متن کاملThree-dimensional custom-designed cementless femoral stem for osteoarthritis secondary to congenital dislocation of the hip.
A clinical and radiological study was conducted on 97 total hip replacements performed for congenital hip dislocation in 79 patients between 1989 and 1998 using a three-dimensional custom-made cementless stem. The mean age at operation was 48 years (17 to 72) and the mean follow-up was for 123 months (83 to 182). According to the Crowe classification, there were 37 class I, 28 class II, 13 clas...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 91 3 شماره
صفحات -
تاریخ انتشار 2009