Cardioverter-defibrillator box damage secondary to lead insulation defect.

نویسندگان

  • Venkata S Dontaraju
  • Ian Loke
  • John D Skehan
چکیده

A 39-year-old Caucasian female with hypertrophic obstructive cardiomyopathy and with history of syncope and other high-risk features for sudden cardiac death underwent implantation of St Jude Photon DR dual chamber cardioverter-defibrillator (ICD). Two years later she was found to be in atrial fibrillation and anticoagulated with warfarin. Three years after the ICD was implanted, it delivered an inappropriate single shock and was found to be in hardware reset mode with no defibrillator capacity. The device was explanted and sent to St Jude Medical for investigation. A new Medtronic Intrinsic ICD was implanted uneventfully and lead tested at device change. Defibrillation thresholds were deferred as she was in atrial fibrillation and not adequately anticoagulated. Few days later she collapsed and died. Device interrogation revealed ventricular tachycardia degenerated in to ventricular fibrillation and several unsuccessful shocks delivered. Investigation of St Jude’s device revealed a dent and an arc mark on the device box and Spectrography revealed the presence of lead material. It is believed that insulation of the high voltage lead was damaged and caused the lead to arc to the box during the high voltage therapy resulting in the damage to the box (dent and arc) and failure of energy delivery to heart. No defibrillator output was traced to shorted transistor. The battery longevity was normal. So two ICDs from different manufacturers are damaged due to lead insulation defect. Along with other parameters, ICD boxes should be inspected for any obvious damage during box change.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation

A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead...

متن کامل

Developing a crush: acute implantable cardioverter-defibrillator lead insulation break in a patient with multiple leads.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Developing a crush: acute implantable cardioverter-...

متن کامل

THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL

Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...

متن کامل

The Subcutaneous Implantable Cardioverter-Defibrillator: Not For All

Implantable cardioverter-defibrillators (ICDs) have a proven role in primary and secondary prevention of sudden death from ventricular arrhythmias. Transvenous ICD systems despite this proven benefit, have been plagued by lead problems including fracture, insulation breaks and recalls; and issues with transvenous access including pneumothorax and venous obstruction. Infected or fractured leads ...

متن کامل

Failure of a novel silicone–polyurethane copolymer (Optim™) to prevent implantable cardioverter-defibrillator lead insulation abrasions

AIM The purpose of this study was to determine if Optim™, a unique copolymer of silicone and polyurethane, protects Riata ST Optim and Durata implantable cardioverter-defibrillator (ICD) leads (SJM, St Jude Medical Inc., Sylmar, CA, USA) from abrasions that cause lead failure. METHODS AND RESULTS We searched the US Food and Drug Administration's (FDA's) Manufacturers and User Device Experienc...

متن کامل