Measurements of Intraocular Pressure with Tono-Pen and Goldmann Applanation Tonometers in Subjects with Systemic Illness
نویسندگان
چکیده
Objective: The purpose of this preliminary study was to compare differences in the measurement of intraocular pressure (IOP) between TonoPen tonometry and Goldmann applanation tonometry (GAT) in healthy individuals and those with coexisting illnesses. Patients and Methods: This prospective study included 42 individuals who were further categorized into two groups. Group 1 consisted of 20 healthy individuals and group 2 contained 22 participants who had various systemic illnesses such as congestive heart failure, chronic bronchitis, uremia, liver cirrhosis and cerebral ischemia stroke. One eye of each of the 42 participants was randomly selected and the IOP was measured with the Tono-Pen and GAT. In addition, the central corneal thickness (CCT), axial length (AL) and keratometric power (K) were measured in the same eyeball of each participant. Results: There was no significant difference in the IOP values between the Tono-Pen and GAT in these two groups (p > 0.05). The average IOP values in the healthy individuals were 14.5 ± 3.5 and 13.7 ± 3.4 mmHg with the Tono-Pen and GAT, respectively. The IOP values in the subjects with coexisting systemic illnesses were 13.8 ± 2.9 and 13.1 ± 3.2 mmHg with the Tono-Pen and GAT, respectively. The IOP values were positively correlated with CCT with both the Tono-Pen (group 1, p = 0.042; group 2, p = 0.039) and GAT (group 1, p = 0.020; group 2, p = 0.015). There were no significant differences in the age of subjects, AL or K (p > 0.05) in these two groups. Conclusion: In our preliminary study, the IOP values were positively correlated with the CCT with Tono-Pen tonometry and GAT in these two groups. The IOP values were not significantly different between the Tono-Pen and GAT in the subjects with systemic illness. [Tzu Chi Med J 2007;19(4):241–244] Article info Article history: Received: April 19, 2007 Revised: May 19, 2007 Accepted: June 15, 2007
منابع مشابه
Intraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea
Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...
متن کاملClinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.
A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...
متن کاملComparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL
OBJECTIVE WE AIMED TO COMPARE INTRAOCULAR PRESSURE (IOP) MEASUREMENTS VIA THREE DIFFERENT TONOMETERS: the Goldmann applanation tonometer (GAT), the Tono-Pen(®) XL (TPXL), and a non-contact airpuff tonometer (NCT). METHODS This was a cross-sectional study of 200 eyes from 200 patients. Right eyes of all patients were included in this study. IOP was measured via GAT, NCT, and TPXL by three phys...
متن کاملDynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies
Unspecified Posted at the Zurich Open Repository and Archive, University of Zurich ZORA URL: http://doi.org/10.5167/uzh-101903 Published Version Originally published at: Neuburger, M; Grosswendt, J; Lautebach, S; Maier, P; Birnbaum, F; Böhringer, D; Funk, J; Jordan, J F; Reinhard, T (2014). Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral...
متن کاملTono-Pen tonometry in normal and in post-keratoplasty eyes.
Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 82 eyes of 82 patients with normal corneas and in 54 eyes of 54 patients who had undergone penetrating keratoplasty and whose corneas did not preclude the use of Goldmann tonometer. We found that the intraocular pressure (IOP) in 48% of the eyes with normal corneas and in 57% after keratoplasty has different measureme...
متن کامل