Development and validation of a method for mercury determination in seawater for the process control of a candidate certified reference material
نویسندگان
چکیده
A simple, robust and reliable method for mercury determination in seawater matrices based on the combination of cold vapour generation and inductively coupled plasma mass spectrometry (CV-ICP-MS) and its complete in-house validation are described. The method validation covers parameters such as linearity, limit of detection (LOD), limit of quantification (LOQ), trueness, repeatability, intermediate precision and robustness. A calibration curve covering the whole working range was achieved with coefficients of determination typically higher than 0.9992. The repeatability of the method (RSDrep) was 0.5 %, and the intermediate precision was 2.3 % at the target mass fraction of 20 ng/kg. Moreover, the method was robust with respect to the salinity of the seawater. The limit of quantification was 2.7 ng/kg, which corresponds to 13.5 % of the target mass fraction in the future certified reference material (20 ng/kg). An uncertainty budget for the measurement of mercury in seawater has been established. The relative expanded (k = 2) combined uncertainty is 6 %. The performance of the validated method was demonstrated by generating results for process control and a homogeneity study for the production of a candidate certified reference material.
منابع مشابه
Determining Total Mercury in Samples from the Persian Gulf and theCaspian Sea: Comparison of Dry Ash and Wet Extraction Mothods
Monitoring of mercury in environmental samples, with its proven toxicity on the food chain, requires sensitive and accurate analytical techniques. In this study, two methods for identification and quantification of total mercury in biological and sediment samples are compared: i) dry ash preparation and subsequent established procedures for Combustion Atomic Absorption Spectrometry, (AMA 254) ...
متن کاملPreconcentration of Cu(II)with magnetic nanocomposite sorbentsprior to determination by Flame Atomic Absorption Spectrometer
A simple, highly sensitive, accurate and selective method for determination of trace amounts of Cu2+ in water samples .In this paper, Chitosan was used to encapsulate nano-Fe3O4 and produce a nano-Fe3O4-CHITOSAN sorbent based new sorbent was prepared. This was treated with ethylenediamine (ED) in another solvent-free procedure for the formation of a novel nano-Fe3O4-Chitosan-ED.Flame atomic abs...
متن کاملPreconcentration of Cu(II)with magnetic nanocomposite sorbentsprior to determination by Flame Atomic Absorption Spectrometer
A simple, highly sensitive, accurate and selective method for determination of trace amounts of Cu2+ in water samples .In this paper, Chitosan was used to encapsulate nano-Fe3O4 and produce a nano-Fe3O4-CHITOSAN sorbent based new sorbent was prepared. This was treated with ethylenediamine (ED) in another solvent-free procedure for the formation of a novel nano-Fe3O4-Chitosan-ED.Flame atomic abs...
متن کاملDevelopment and Validation of a HPLC Method for Determination of Pefloxacin in Tablet and Human Plasma
Objective(s) Developing and validating a simple, efficient, reproducible and economic reversed phase high performance liquid chromatographic (RP-HPLC) method for the quantitative determination of pefloxacin in bulk material, tablets and in human plasma. Materials and Methods A shim-pack CLC-ODS column and a mobile phase constituting acetonitrile: 0.025 M phosphoric acid solution (13:87 v/v, ...
متن کاملValidation of a genus-specific gene; TPS, used as internal control in quantitative Real Time PCR of transgenic cotton
Identification of genes with invariant levels of gene expression is a prerequisite for validating transcriptomic changes accompanying development. Ideally expression of these genes should be independent of the morphogenetic process or environmental condition.We report here the validation of internal control gene i.e.TPS (trehalose 6-phosphate-synthase) in cotton (Gossypium spp), using TaqMan sy...
متن کامل