Validation of the BPLab® 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol

نویسندگان

  • Sergey A Koudryavtcev
  • Vyacheslav M Lazarev
چکیده

INTRODUCTION Automatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab(®) ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice. METHODS A group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique were recruited for this study. The results of the BP measurement obtained by a qualified operator using the BPLab(®) device were compared with the BP values measured using the Korotkov auscultatory method. Data were obtained simultaneously by two experts with experience of over 10 years and had completed a noninvasive BP measurement standardization training course. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were analyzed according to the criteria specified in the BHS-93 protocol. RESULTS The device passed the requirements of the European Standard BS EN 1060-4:2004 and was graded 'A' according to the criteria of the BHS protocol for both systolic BP and diastolic BP. CONCLUSION The BPLab(®) 24-hour ambulatory blood pressure monitoring device may be recommended for extensive clinical use.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the BPLab® sphygmomanometer for ambulatory blood pressure monitoring with mercury sphygmomanometry in pregnant women: validation study according to the British Hypertension Society protocol

The purpose of this study was to validate the automated BPLab® sphygmomanometer for ambulatory blood pressure monitoring (ABPM) in pregnant women according to Part II of the 1993 British Hypertension Society protocol. Pregnant women attending the antenatal clinic were randomly asked to participate (n=30). The BPLab sphygmomanometer was tested on pregnant women in this study and achieved A/A rat...

متن کامل

Validation of the BPLab® 24-hour blood pressure monitoring system in a pediatric population according to the 1993 British Hypertension Society protocol

BACKGROUND Automatic 24-hour ambulatory blood pressure (BP) monitoring (ABPM) is a basic procedure performed in adults with arterial hypertension, but ABPM monitors have become widely used in pediatric practice only recently. The main problem is the lack of common normative data sets for ABPM in children and the small number of appropriate monitors that can be used for analysis of the 24-hour B...

متن کامل

Twenty-four-hour ambulatory blood pressure monitoring: a review of validation data.

As the clinical applications for 24-h ambulatory blood pressure monitoring expand, market demands increase, and there are now at least 13 manufacturers producing elaborate and expensive systems for recording 24-h ambulatory pressures. It is often difficult to assess the accuracy and performance characteristics of these devices and failure to standardize validation makes it difficult to compare ...

متن کامل

Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol.

OBJECTIVES We tested the accuracy of the UA-705 blood pressure semi-automatic monitor. METHODS Device evaluation was performed according to the modified British Hypertension Society protocol released in 1993. Eighty-five patients with characteristics outlined in the British Hypertension Society protocol were recruited among those attending our out-patient clinic. Sequential readings were take...

متن کامل

Proposals for simplifying the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation.

Correspondence to Professor Eoin O’Brien, The Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland. Tel : +353 1 809 21qO; fax: i-353 1 809 3005; e-mail: [email protected] Introduction Experience with the protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI) has provided valuable insight into the methodological problems a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2011