Distortion Correction in EPI Using an Extended PSF Method with a Reversed Phase Gradient Approach
نویسندگان
چکیده
In echo-planar imaging (EPI), such as commonly used for functional MRI (fMRI) and diffusion-tensor imaging (DTI), compressed distortion is a more difficult challenge than local stretching as spatial information can be lost in strongly compressed areas. In addition, the effects are more severe at ultra-high field (UHF) such as 7T due to increased field inhomogeneity. To resolve this problem, two EPIs with opposite phase-encoding (PE) polarity were acquired and combined after distortion correction. For distortion correction, a point spread function (PSF) mapping method was chosen due to its high correction accuracy and extended to perform distortion correction of both EPIs with opposite PE polarity thus reducing the PSF reference scan time. Because the amount of spatial information differs between the opposite PE datasets, the method was further extended to incorporate a weighted combination of the two distortion-corrected images to maximize the spatial information content of a final corrected image. The correction accuracy of the proposed method was evaluated in distortion-corrected data using both forward and reverse phase-encoded PSF reference data and compared with the reversed gradient approaches suggested previously. Further we demonstrate that the extended PSF method with an improved weighted combination can recover local distortions and spatial information loss and be applied successfully not only to spin-echo EPI, but also to gradient-echo EPIs acquired with both PE directions to perform geometrically accurate image reconstruction.
منابع مشابه
Correction of geometric distortion in Propeller echo planar imaging using a modified reversed gradient approach.
OBJECTIVE This study investigates the application of a modified reversed gradient algorithm to the Propeller-EPI imaging method (periodically rotated overlapping parallel lines with enhanced reconstruction based on echo-planar imaging readout) for corrections of geometric distortions due to the EPI readout. MATERIALS AND METHODS Propeller-EPI acquisition was executed with 360-degree rotationa...
متن کاملImplementation and Application of PSF-Based EPI Distortion Correction to High Field Animal Imaging
The purpose of this work is to demonstrate the functionality and performance of a PSF-based geometric distortion correction for high-field functional animal EPI. The EPI method was extended to measure the PSF and a postprocessing chain was implemented in Matlab for offline distortion correction. The correction procedure was applied to phantom and in vivo imaging of mice and rats at 9.4T using d...
متن کاملImage distortion correction in EPI: comparison of field mapping with point spread function mapping.
Echo-planar imaging (EPI) can provide rapid imaging by acquiring a complete k-space data set in a single acquisition. However, this approach suffers from distortion effects in geometry and intensity, resulting in poor image quality. The distortions, caused primarily by field inhomogeneities, lead to intensity loss and voxel shifts, the latter of which are particularly severe in the phase-encode...
متن کاملDistortion correction in spinal cord DTI: What’s the best approach?
Introduction Susceptibility artifacts induced by B0-field inhomogeneities are common issue in echo planar imaging (EPI). They are particularly problematic in the spinal cord region, inducing large geometric and intensity distortions [1]. Although non-EPI techniques could be used instead [2], ultra fast imaging is sometimes a requirement in protocols such as in high angular resolution diffusion ...
متن کاملJacobian weighting of distortion corrected EPI data
Introduction: The Reversed Gradient Polarity Method (‘RGPM’) has proven effective in correcting for geometric distortions in EPI data from two oppositely distorted magnitude EPI images, in particular when the estimation of smooth displacement fields are enforced via basis functions and optionally combined with simultaneous motion correction. This method does not involve extensive calibration me...
متن کامل