Effect of inherent misalignment and head motion in neurological PET/MR with the Philips Ingenuity TF – phantom and patient study
نویسندگان
چکیده
Turku PET Centre, Turku University Hospital, Turku, Finland The aim of the study was to evaluate the effect of misalignment and head motion on image quantification in PET/MR with a novel brain phantom and a healthy control group. The phantom was imaged at two time points in PET/MR, concurrently with PET and PET/CT. Phantom images were evaluated visually and the relative difference in hemispheric accumulation was calculated. Difference in cortical accumulation in a healthy control group was evaluated from non-attenuation corrected (NAC) and MR attenuation corrected (MRAC) images. Regional ROI mean values from F-FDG ratio images and regional hemispheric asymmetries were calculated. Controls were divided to high and low asymmetry groups. A student’s t-test (p<0.005) for group difference and NAC versus MRAC data was performed. Finally, mean PET-MR registration parameters were measured. Only the first phantom scan exhibited asymmetry in lateral frontal cortex (17%) and temporal cortex (19%). Correcting the misalignment of 2.63mm reduced the asymmetry to less than 5%, to a level seen in PET and PET/CT. A significant asymmetry was found in the temporal and parietal cortex between groups in MRAC data with no significant asymmetry in NAC data. Asymmetries in affected MRAC data in temporal and parietal cortex were 9.4% and 11.7%. NAC data from both groups had asymmetry less than 5% in all regions. Both groups had significant yand z-translation, while only the asymmetry group had significant z-rotation and x-translation. The shift in x-, y-, or z-direction in both groups was less than 4 mm, with no significant differences. Thus, PET-MRAC misalignment may cause underand overestimation of attenuation in the lines of response on opposite sides of the cortical regions, resulting to asymmetric difference between the hemispheres. Our findings stress the need for novel QC procedures for PET-MR alignment and suggest confirming the quality of PET-MRAC alignment from PET-NAC images.
منابع مشابه
Long-term stability of the MR system of the Philips Ingenuity TF.
Turku PET Center, Turku, Finland The Philips Ingenuity TF is a sequential PET/MR with an Achieva 3T X-series MR and Gemini TF PET system connected with a rotating table. Both the PET and MR systems are designed to minimize mutual system interference. However, a longitudinal study concerning the stability of the MR system has not been conducted before. The stability of the MR system affects diff...
متن کاملPreliminary evaluation of the MLAA algorithm with the Philips Ingenuity PET/MR
PET Center, Institute of Radiopharmaceutical Cancer Research, Helmholtz-Zentrum Dresden-Rossendorf, Bautzner Landstrae 400, 01328 Dresden, Germany Combined PET/MR is a promising tool for simultaneous investigation of soft tissue morphology and function. However, contrary to CT, MR images do not provide information on photon attenuation in tissue. In the currently available systems issue is solv...
متن کاملComparison of different tube-of-response (TOR) models for resolution recovery in PET image reconstruction for the Philips Ingenuity TF PET/MR
Helmholtz-Center DresdenRossendorf, Institute of Radiopharmaceutical Cancer Research, PET Center, Dresden, Germany Recently, we have proposed a method for on-the-fly system matrix computation where the tube-of-response (TOR) is approximated as a cylinder with constant density (TORCD) and the cubic voxels are replaced by spheres. We could show that with this model the PET image quality can be no...
متن کاملRespiratory motion correction in prostate cancer positron emission tomography: A study on patients and phantom simulation
Introduction: To investigate the effects of breathing cycle and tree diaphragm motions on prostate cancer tumors standard uptake value (SUV) during positron emission tomography (PET) and to correct it. Materials and methods: Respiratory motion traces were simulated on the common patient breathing cycle and tree diaphragm motio...
متن کاملPartial volume correction of brain PET studies using iterative deconvolution in combination with HYPR denoising
BACKGROUND Accurate quantification of PET studies depends on the spatial resolution of the PET data. The commonly limited PET resolution results in partial volume effects (PVE). Iterative deconvolution methods (IDM) have been proposed as a means to correct for PVE. IDM improves spatial resolution of PET studies without the need for structural information (e.g. MR scans). On the other hand, deco...