Development of Ultrafast Laser-based Micro-ct System for Small Animal Imaging
نویسندگان
چکیده
We investigated performance of ultrafast (30 fs –200 fs) laser-based x-ray source as an alternative to a microfocal x-ray tube used in micro-CT systems. A number of solid targets including Ge, Mo, Rh, Ag, Sn, Ba, La, Nd with matching filters was used. We optimized conditions for x-rays generation, measured x-ray spectra and obtained images of small animals using various characteristic x-ray lines and various imaging geometries. X-ray spectra produced by ultrafast laser are advantageous for micro-CT imaging because most of the emission is in narrow characteristic lines. The spectra can be rapidly changed, via target change, and easily matched to the imaging task. X-ray fluence can exceed fluence produced by conventional microfocal tube. Effective x-ray focal spot size is about 5 μm. Good quality projection images of small animals were obtained: (a) in a single energy absorption mode with single x-ray line matching animal thickness and density, and (b) in the dual-energy absorption mode with two x-ray lines bracketing absorption energy of the I-based and Ba-based contrast agent, respectively, and (c) in the in line phase-contrast enhancing geometry using single x-ray line matching animal thickness and density. We conclude that the ultrafast laser-based, compact, x-ray source is a promising alternative to a microfocal x-ray tube in micro-CT systems. Its utilization might result in faster scans with lower radiation dose, and with better spatial and contrast resolution. In addition, it might allow practical implementation of dual-energy and phase-contrast imaging micro-CT that not possible with conventional micro-CT.
منابع مشابه
Design of Small Animal Computed Tomography Imaging for in vitro and in vivo Studies
Introduction: Mini Computed Tomography (mini-CT) was suggested in biomedical research to investigate tissues and small animals. We present designed and built a mini x-ray computed tomography (mini-CT) for small animals as well as industrial component imaging. Materials and Methods: The system used in this study includes a X-ray tube 20kV to 160kV and a flat pa...
متن کاملInfluence of Interface Thermal Resistance on Relaxation Dynamics of Metal-Dielectric Nanocomposite Materials under Ultrafast Pulse Laser Excitation
Nanocomposite materials, including noble metal nanoparticles embedded in a dielectric host medium, are interesting because of their optical properties linked to surface plasmon resonance phenomena. For studding of nonlinear optical properties and/or energy transfer process, these materials may be excited by ultrashort pulse laser with a temporal width varying from some femtoseconds to some hund...
متن کاملOptimization of Imaging Parameters in Micro-CT Scanner Based On Signal-To-Noise Ratio for the Analysis of Urinary Stone Composition
Introduction: Micro-CT scanner with a resolution of about 5 micrometers is one of the modalities used to create three-dimensional/two-dimensional images of urinary stones. This study aimed to optimize imaging parameters in micro-computed tomography (CT) scanner based on the signal-to-noise ratio (SNR) of urinary stones for the analysis of stone composition. <stro...
متن کاملGraphene Quantum Dots-Coated Bismuth Nanoparticles for X-ray CT Imaging-Guided Photothermal therapy of Cancer Cells
Introduction: Theranostic nanoparticles, which integrate both diagnostic and therapeutic capabilities into one nanoagent, has great promise to ablate more effective tumoral tissue by optimizing and real-time monitoring treatment interventions, as well as monitoring therapeutic response to corresponding effect. Multifunctional theranostic nanoagent based on graphene quantum dots...
متن کاملIn vivo small animal micro-CT using nanoparticle contrast agents
Computed tomography (CT) is one of the most valuable modalities for in vivo imaging because it is fast, high-resolution, cost-effective, and non-invasive. Moreover, CT is heavily used not only in the clinic (for both diagnostics and treatment planning) but also in preclinical research as micro-CT. Although CT is inherently effective for lung and bone imaging, soft tissue imaging requires the us...
متن کامل