A quality assurance method with submillimeter accuracy for stereotactic linear accelerators

نویسندگان

  • Jimm Grimm
  • Shu‐Ya Lisa Grimm
  • Indra J. Das
  • Yunping Zhu
  • Inhwan Yeo
  • Jinyu Xue
  • Larry Simpson
  • Dayee Jacob
  • Abhirup Sarkar
چکیده

The Stereotactic Alignment for Linear Accelerator (S. A. Linac) system is developed to conveniently improve the alignment accuracy of a conventional linac equipped with stereotactic cones. From the Winston-Lutz test, the SAlinac system performs three-dimensional (3D) reconstruction of the quality assurance (QA) ball coordinates with respect to the radiation isocenter, and combines this information with digital images of the laser target to determine the absolute position of the room lasers. A handheld device provides near-real-time repositioning advice to enable the user to align the QA ball and room lasers to within 0.25 mm of the centroid of the radiation isocenter. The results of 37 Winston-Lutz tests over 68 days showed that the median 3D QA ball alignment error was 0.09 mm, and 97% of the time the 3D error was ≤ 0.25 mm. All 3D isocentric errors in the study were 0.3 mm or less. The median x and y laser alignment coordinate error was 0.09 mm, and 94% of the time the x and y laser error was ≤ 0.25 mm. A phantom test showed that the system can make submillimeter end-to-end accuracy achievable, making a conventional linac a "Submillimeter Knife".

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

New Method of Quality Control Test for Light and Radiation Field Coincidence in Medical Linear Accelerators

Introduction: The evaluation of X-ray and light field coincidence in linear accelerators as a quality control test is often performed subjectively, involving the manual marking of films and their visual inspection following the irradiation. Therefore, the present study aimed to develop an objective method for the performance of this test leading to the increased levels...

متن کامل

New electrical device for field coincidence quality control test in linear accelerators

Testing of the coincidence of a linear accelerators X-ray field and the light field as a quality control test is often done in a subjective method, involving the manual marking of pieces of film and the visual inspection of the film after irradiation. The purpose of this study was to develop an objective method for performing this test, while also increasing the accuracy, precision and the spee...

متن کامل

Quality assurance using a photodiode array

Improved treatment techniques in radiation therapy provide incentive to reduce treatment margins, thereby increasing the necessity for more accurate geometrical setup of the linear accelerator and accompanying components. In the present paper, we describe the development of a novel device that enables precise and automated measurement of geometric parameters for the purpose of improving initial...

متن کامل

Beam focal spot position: The forgotten linac QA parameter. An EPID‐based phantomless method for routine Stereotactic linac QA

Modern day Stereotactic treatments require high geometric accuracy of the delivered treatment. To achieve the required accuracy the IGRT imaging isocenter needs to closely coincide with the treatment beam isocenter. An influence on this isocenter coincidence and on the spatial positioning of the beam itself is the alignment of the treatment beam focal spot with collimator rotation axis. The pos...

متن کامل

The Quality Control of Intensity Modulated Radiation Therapy (IMRT) for ONCOR Siemens Linear Accelerators Using Film Dosimetry

Introduction Intensity Modulated Radiation Therapy (IMRT) has made a significant progress in radiation therapy centers in recent years. In this method, each radiation beam is divided into many subfields that create a field with a modulated intensity. Considering the complexity of this method, the quality control for IMRT is a topic of interest for researchers. This article is about the various ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2010