Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
نویسندگان
چکیده
This work investigated the suitability of passive dosimeters for reference dosimetry in small fields with acceptable accuracy. Absorbed dose to water rate was determined in nine small radiation fields with diameters between 4 and 35 mm in a Leksell Gamma Knife (LGK) and a modified linear accelerator (linac) for stereotactic radiosurgery treatments. Measurements were made using Gafchromic film (MD-V2-55), alanine and thermoluminescent (TLD-100) dosimeters and compared with conventional dosimetry systems. Detectors were calibrated in terms of absorbed dose to water in (60)Co gamma-ray and 6 MV x-ray reference (10×10 cm(2)) fields using an ionization chamber calibrated at a standards laboratory. Absorbed dose to water rate computed with MD-V2-55 was higher than that obtained with the others dosimeters, possibly due to a smaller volume averaging effect. Ratio between the dose-rates determined with each dosimeter and those obtained with the film was evaluated for both treatment modalities. For the LGK, the ratio decreased as the dosimeter size increased and remained constant for collimator diameters larger than 8 mm. The same behaviour was observed for the linac and the ratio increased with field size, independent of the dosimeter used. These behaviours could be explained as an averaging volume effect due to dose gradient and lack of electronic equilibrium. Evaluation of the output factors for the LGK collimators indicated that, even when agreement was observed between Monte Carlo simulation and measurements with different dosimeters, this does not warrant that the absorbed dose to water rate in the field was properly known and thus, investigation of the reference dosimetry should be an important issue. These results indicated that alanine dosimeter provides a high degree of accuracy but cannot be used in fields smaller than 20 mm diameter. Gafchromic film can be considered as a suitable methodology for reference dosimetry. TLD dosimeters are not appropriate in fields smaller than 10 mm diameters.
منابع مشابه
Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملImpact of region of interest size and location in Gafchromic film dosimetry
Introduction: Accurate film dosimetry requires careful consideration of sources of uncertainty. Some of the sources of uncertainty are dependent on the size and location of region of interest (ROI), especially in small fields. Avoiding the penumbra is often a reason for using a small ROI. In contrast, choosing very small ROIs may increase uncertainty due to the reduction of th...
متن کاملEvaluation of MRI-based MAGIC polymer gel dosimeter in small photon fields
Background: Accurate small radiation field dosimetry is essential in modern radiotherapy techniques such as stereotactic radiosurgery (SRS) and intensity modulated radiotherapy (IMRT). Precise measurement of dosimetric parameters such as beam profile, percentage depth doses and output factor of these beams are complicated due to the electron disequilibrium and the steep dose gradients. In the p...
متن کاملMeasurement Of Scatter Factors For Small Photon Fields Using Gaf chromic EBT2 Film
Introduction: Small field dosimetry is challenging for radiotherapy dosimetry. measurement of the output factor in the air and water (Sc, Scp) is one of the input parameters for commissioning of treatment planning systems and beam modeling. The aims of this study are to measured Sc,Scp for small fields with EBT2 and Ion chamber and design a appropriate mini-phantom for small fi...
متن کاملRadiographic film dosimetry for IMRT fields in the near‐surface buildup region
Radiographic film dosimetry provides fast, convenient 2-D dose distributions, but is challenged by the dependence of film response on scatter conditions (i.e., energy dependence). Verification of delivered dose in the surface buildup region is important for intensity modulated radiation therapy (IMRT) when volumes of interest encroach on these regions (e.g., head/neck, breast). The current work...