Variability in aerosol output of the DeVilbiss 646 jet nebulizer.

نویسندگان

  • R Jörres
  • D Nowak
  • K Rabe
  • H Magnussen
چکیده

Ten new jet nebulizers (DeVilbiss 646) were studied to determine their consistency of output. Each nebulizer, containing 2 ml of saline solution, was run for 1 min in triplicate at four different straw (capillary tube) positions, keeping all other variables constant. Total output in milliliters per minute and volume of aerosol in the respirable range (1.0 to 5.0 microns) were measured. There was significant variability in output and volume of output in the respirable range (VORR) for every nebulizer at each of the tested straw positions (p less than 0.0001). Irregularities in manufacture of the bowl and straw appear to be responsible for this significant variability. In summary, there is considerable intranebulizer and internebulizer variability that could influence both patient care and medical use for diagnostic purposes.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measuring nebulizer output. Aerosol production vs gravimetric analysis.

STUDY OBJECTIVES The function of jet nebulizers has been measured traditionally by gravimetric methods, i.e., by weighing nebulizers before and after nebulization. Newer techniques measure aerosol output directly by analyzing aerosolized drug or tracer, i.e., radioactive 99mTc. Because of evaporation, the equivalence of these methods is uncertain. The aim of this study was to determine if the g...

متن کامل

Output characteristics of DeVilbiss No. 40 hand-held jet nebulizers.

DeVilbiss No. 40 hand-held nebulizers are widely used for quantifying airway responsiveness in large populations using pharmacological agents. We examined the aerosol characteristics of five nebulizers. Within each device, the aerosol output and droplet size were reasonably stable over a wide range of bulb pressures, although there were considerable differences in output characteristics between...

متن کامل

Methods of testing nonspecific bronchial hyperresponsiveness.

Many methods of testing nonspecific bronchial responsiveness have been devised, since Curry first demonstrated the heightened bronchial responsiveness of asthmatics to histamine. The preferred rou1e of administration of bronchoconstrictors at the presem Lime is by inhalation, rather than systemically, in order to minimize unacceptable side effects. The agents inhaled have included irritants, ch...

متن کامل

A comparison of commercial jet nebulizers.

Seventeen commercially available jet nebulizers from 15 commercial sources were studied (Acorn-I, Acorn-II, AquaTower, AVA-NEB, Cirrhus, Dart, DeVilbiss 646, Downdraft, Fan Jet, MB-5, Misty Neb, PARI LC JET, PARI-JET, Salter 8900, Sidestream, Updraft-II, Whisper Jet). All nebulizers were filled with 2 ml of saline solution plus 0.5 ml of albuterol and powered with the same source (DeVilbiss Pul...

متن کامل

Differences in aerosol output and airways responsiveness between the DeVilbiss 40 and 45 hand held nebulisers.

BACKGROUND The DeVilbiss 40 glass hand held nebulisers have been widely used for airways responsiveness testing in epidemiological surveys of asthma. These nebulisers have been superseded in some recent studies by the DeVilbiss 45 plastic hand held nebulisers with the assumption that they are interchangeable. This study compared the aerosol outputs of the DeVilbiss 40 and DeVilbiss 45 nebuliser...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Chest

دوره 102 5  شماره 

صفحات  -