WE-G-BRB-09: CyberKnife Patient Specific QA Using a 4D Cylindrical Diode Array System.

نویسندگان

  • S Koren
  • I Veltchev
  • J Fan
  • M Lin
  • A Guemnie Tafo
  • C Ma
چکیده

PURPOSE To examine and facilitate the feasibility of the ArcCheck cylindrical diode array system as a patient specific QA device for CyberKnife radiosurgery delivery. METHODS There is an obvious necessity for CyberKnife robotic radiosurgery patient QA procedures for hypofractionated treatment of larger planned treatment volumes (PTV), e.g. prostate. This need will increase when the future CyberKnife MLC is introduced. The small unflattened CyberKnife fields, along with the variation of beam-to-detector spatial angles, pose a significant detection challenge for dosimetric systems. The feasibility of the ArcCheck (Sun Nuclear Inc.) cylindrical diode array system for patient-specific QA on the CyberKnife is demonstrated using a beam-to-diode specific angular correction that was developed and has been applied. For localization and tracking, four gold seed fiducial markers were embedded in the system's central plug. We used a Monte Carlo 1% uncertainty for the dose calculation. RESULTS By disabling the Linac based corrections and applying the custom CyberKnife correction that we developed, the passing rate increased from 39.6% to 99.8% using a 3%3mm gamma criteria for a given lung case. An additional lung case passed 98.5%. In both cases, a 10% dose threshold was used. In addition, brain, trigeminal nerve and lung cases with synchrony tracking are being investigated. CONCLUSIONS We demonstrated the ArcCheck feasibility for CyberKnife patient specific QA performance. The custom CK angular correction that we developed and applied showed a high passing rate for the lung cases. A verification of the polar angle response should be conducted, in addition to the azimuthal angle that was verified for Linacs. Any data that is being retrieved is additional data to the current chamber point measurement procedures.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Robotic radiosurgery system patient‐specific QA for extracranial treatments using the planar ion chamber array and the cylindrical diode array

Robotic radiosurgery system has been increasingly employed for extracranial treatments. This work is aimed to study the feasibility of a cylindrical diode array and a planar ion chamber array for patient-specific QA with this robotic radiosurgery system and compare their performance. Fiducial markers were implanted in both systems to enable image-based setup. An in-house program was developed t...

متن کامل

2D linear array device as a quality assurance tool in brachytherapy applications

Background: External beam radiotherapy and brachytherapy plays a vital role in the management of cancer cervix.  High dose rate brachytherapy is being presently used worldwide for the brachytherapy applications. At present, 2-Dimensional linear array detectors   are the most common QA tool used for pretreatment patient specific quality assurance in external beam radiotherapy alon...

متن کامل

On the sensitivity of patient‐specific IMRT QA to MLC positioning errors

Accurate multileaf collimator (MLC) leaf positioning plays an essential role in the effective implementation of intensity modulated radiation therapy (IMRT). This work evaluates the sensitivity of current patient-specific IMRT quality assurance (QA) procedures to minor MLC leaf positioning errors. Random errors of up to 2 mm and systematic errors of +/-1 mm and +/-2 mm in MLC leaf positions wer...

متن کامل

3D Analysis of Intensity-Modulated Radiation Therapy Quality Assurance Measurement using a 2D Diode Array

Intensity-modulated radiation therapy (IMRT) quality assurance (QA) is often performed using a 2D device and compares measured and computed fluence maps to determine if a field passes or fails certain dose and position criteria. The effects of a measured deviation to the 3D patient spatial dosimetry and dose volume histogram (DVH) are largely unknown because they cannot be analyzed using commer...

متن کامل

Effects on the photon beam from an electromagnetic array used for patient localization and tumor tracking

One of the main components in a Calypso 4D localization and tracking system is an electromagnetic array placed above patients that is used for target monitoring during radiation treatment. The beam attenuation and beam spoiling properties of the Calypso electromagnetic array at various beam angles were investigated. Measurements were performed on a Varian Clinac iX linear accelerator with 6 MV ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part28  شماره 

صفحات  -

تاریخ انتشار 2012