Inhalation of single vs multiple metered-dose bronchodilator actuations from reservoir devices. An in vitro study.
نویسندگان
چکیده
UNLABELLED Differences in inhalation technique with reservoir or spacer devices may affect metered-dose inhaler (MDI) dose availability to a patient. PURPOSE This study examined the effect of single vs multiple actuations of an MDI into reservoir devices on dose delivery of albuterol, with three clinically available reservoir brands. METHODS An in vitro lung model simulated inspiration from the MDI reservoir system. Albuterol (Proventil; Schering) was delivered by MDI, with the Monaghan Aerochamber, the Diemolding Healthcare Division (DHD) aerosol cloud enhancer (ACE), and the Schering InspirEase, using standardized volumes and inspiratory flows of 30 L min(-1). The MDI was actuated into each brand of reservoir 1, 2, or 3 times in rapid succession, followed by a single inhalation. Aerosol dose at the reservoir mouthpiece was captured on a cotton filter, dissolved in ethanol, and measured with a spectrophotometer at 278 nm. RESULTS For all three brands of reservoir, less accumulated dose of drug is delivered with multiple actuations than with multiple single actuations each followed by inhalation. The total dose in milligrams increased significantly with two multiple actuations compared with one actuation in the Aerochamber and ACE (p<0.01), but not in the InspirEase (p>0.05). The Aerochamber, ACE, and InspirEase delivered a mean total dose (SD) of 0.0264 mg (0.012), 0.0271 mg (0.007), and 0.0136 mg (0.006), respectively, with one actuation compared to 0.0485 mg (0.011), 0.0453 mg (0.013), and 0.0218 mg (0.009) with two multiple actuations. The increase in total dose with three multiple actuations was not significant compared to two actuations for any of the brands tested (p>0.05). Although total dose increased with multiple actuations, a decline in efficiency was seen with two and three multiple actuations, compared to single actuation. The dose delivered per actuation decreased for the Aerochamber, ACE, and InspirEase from 0.0264 mg (0.012), 0.0271 mg (0.007), and 0.0136 mg (0.006) with one actuation, to 0.0243 mg (0.006), 0.0226 mg (0.006), and 0.0109 mg (0.005), respectively, with two multiple actuations, for losses of 8.0%, 16.6%, and 19.9% in dose per actuation for each brand. A further decline in delivery per actuation to 0.0164 mg (0.001), 0.0184 mg (0.004), and 0.0097 mg (0.005) for the 3 brands, respectively, was found with 3 multiple actuations before inhalation. This was a loss of 37.9%, 32.1%, and 28.7% of the dose per single actuation in each brand. There was no significant difference between the Aerochamber and the ACE in dose availability with 1, 2, or 3 actuations, but both of these brands provided significantly more drug than the InspirEase. CONCLUSION Maximal aerosol bronchodilator from an MDI reservoir was given by single actuations each followed by a breath. Two rapid actuations followed by a breath will give a significant accumulation of dose with some loss when compared to two single actuations each followed by inhalation. Three multiple actuations led to a loss of approximately one third of the drug dose obtainable with three single actuations each followed by inhalation, for all three brands.
منابع مشابه
Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug.
BACKGROUND Aerosols generated from metered dose inhalers may be highly charged. The aim of this study was to determine whether lining the walls of a polycarbonate spacer device with an antistatic agent would result in an increase in drug output. The effects of multiple actuations of drug into the spacer device and increasing residence time of drug within the spacer were also determined. METHO...
متن کاملImproving Asthma Management: The Case for Mandatory Inclusion of Dose Counters on All Rescue Bronchodilators
BACKGROUND Asthma remains a serious global health challenge. Poor control of asthma symptoms is due in part to incorrect use of oral inhaler devices that deliver asthma medications, such as poor inhalation technique or use of a metered dose inhaler (MDI) after the recommended number of doses is expelled. OBJECTIVE To review published research on the potential for patients to overestimate or u...
متن کاملMultiple actuations of salbutamol MDI into a spacer device reduce the amount of drug recovered in the respirable range.
We wanted to determine the amount of salbutamol available in respirable particles from the Volumatic spacer device after one, two or five actuations of a metered-dose inhaler into the spacer prior to analysis. A glass multistage liquid impinger was used to determine particle size. Aerosol was sampled after one, two or five actuations from a metered-dose inhaler into a Volumatic spacer. Each exp...
متن کاملEvaluation of inhalation aids of metered dose inhalers in asthmatic children.
Many patients have incorrect aerosol inhalation technique with metered dose inhalers (MDI). Several inhalation aids have been developed and marketed for these patients. Even those who have correct inhalation technique may benefit from these devices. We compared bronchodilator efficacy of two puffs of albuterol aerosol administered either directly from a MDI mouthpiece, or from an inhalation aid...
متن کاملEffect of electrostatic charge, flow, delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants.
BACKGROUND A study was undertaken to determine the influences of electrostatic charge, flow, delay, and multiple actuations on the in vitro delivery of salbutamol generated by a pressurised metered dose inhaler (pMDI) from small volume spacers used in infants. METHODS Ten actuations from a salbutamol pMDI were drawn at different flow rates after either single or multiple actuations, with or w...