Aferição dos equipamentos de laser de baixa intensidade

نویسندگان

  • Thiago Y. Fukuda
  • Julio F. Jesus
  • Marcio G. Santos
  • Claudio Cazarini Junior
  • Maury M. Tanji
  • Helio Plapler
چکیده

Background: Despite the increase in the use of low-level laser therapy (LLLT), there is still a lack of consensus in the literature regarding how often the equipment must be calibrated. Objective: To evaluate the real average power of LLLT devices in the Greater São Paulo area. Methods: For the evaluation, a LaserCheck power meter designed to calibrate continuous equipment was used. The power meter was programmed with data related to the laser’s wavelength to gauge the real average power being emitted. The LLLT devices were evaluated in two ways: first with the device cooled down and then with the device warmed up for 10 minutes. For each condition, three tests were performed. The laser probe was aligned with the power meter, which provided the real average power being emitted by the LLLT device. All of the data and information related to the laser application were collected with the use of a questionnaire filled in by the supervising therapists. Results: The 60 devices evaluated showed deficit in real average power in the cooled-down and warmedup condition. The statistical analysis (ANOVA) showed a significant decrease (p<0.05) in the real average power measured in relation to the manufacturer’s average power. On average, the most common dose in the clinics was 4 J/cm2, and the most desired effects were healing and anti-inflammatory effects. According to the World Association for Laser Therapy (WALT), 1 to 4 J of final energy are necessary to achieve these effects, however only one device was able to reach the recommended therapeutic window. Conclusion: The LLLT devices showed a deficit in real average power that emphasized a lack of order in the application of this tool. The present study also showed the need for periodical calibration of LLLT equipment and a better technical knowledge of the therapists involved.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Low-level laser therapy and Calendula officinalis in repairing diabetic foot ulcers.

OBJECTIVE To evaluate the effects of low-level laser therapy isolated and associated with Calendula officinalis oil in treating diabetic foot ulcers. METHOD An experimental, randomized, controlled, prospective, interventional clinical case study using a quantitative approach. The sample consisted of 32 diabetic patients of both genders. Participants were randomly divided into four groups. Dop...

متن کامل

Clinical Evaluation of the Effects of Low- Intensity Laser (gaalas) on Wound Healing after Gingivoplasty in Humans Avaliação Clínica Dos Efeitos Do Laser Em Baixa Intensidade (gaalas) Na Cicatrização De Gengivoplastia Em Humanos

RESUMO A CLINICAL EVALUATION OF THE EFFECTS OF LOWINTENSITY LASER (GAALAS) ON WOUND HEALING AFTER GINGIVOPLASTY IN HUMANS AVALIAÇÃO CLÍNICA DOS EFEITOS DO LASER EM BAIXA INTENSIDADE (GAALAS) NA CICATRIZAÇÃO DE GENGIVOPLASTIA EM HUMANOS Carla Andreotti DAMANTE1, Sebastião Luiz Aguiar GREGHI2, Adriana Campos Passanezi SANT’ANA2, Euloir PASSANEZI3 1MSc in Periodontology by Bauru Dental School – Un...

متن کامل

[Breast cancer surgery: an old-fashioned procedure for a rising public health problem].

O câncer foi, por muito tempo, considerado um problema de saúde em países desen volvidos. Segundo levantamento de dados atuais e perspectivas futuras, os países em desen volvimento concentram agora a maior parte da população mundial com a doença. No entanto, os sistemas de saúde nessas áreas são reconhecidamente mal estruturados para absorver esse problema emergente. O aumento progressivo na pr...

متن کامل

Effect of the exercise of walkers performed in public squares with spontaneous or prescribed intensity on post-exercise hypotension

OBJECTIVE To quantify the intensity adopted by walkers in public squares and check the occurrence and magnitude of post-exercise hypotension in the spontaneously adopted intensity and in a prescribed intensity. METHODS In 98 volunteers (38 of them being hypertensive), walkers in public squares of the city of João Pessoa, State of Paraíba, Brazil, we have identified the intensity of a usual tr...

متن کامل

Migration of Phthalate-based Plasticizers from PVC and non-PVC Containers and Medical Devices

A presença de cinco plastificantes derivados de ftalatos em equipamentos para administração intravenosa de soluções e em hemodialisadores (HD), que são feitos tanto de policloreto de vinila (PVC) quanto de outros materiais poliméricos, foi investigada. Destes ftalatos, o único autorizado por agências reguladoras é o di-(2-etilhexil) ftalato (DEHP) em equipamentos de PVC. A determinação foi feit...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010