[Investigation of quality control and average glandular dose and image quality in digital mammography in Hokkaido].

نویسندگان

  • Kunio Kurowarabi
  • Hiroko Abe
  • Hiroshi Horita
  • Kazuyuki Kaneta
چکیده

A questionnaire survey about mammography in Hokkaido was mailed to 121 facilities from August to September 2009. We surveyed the conditions of digital mammography with regard to quality control (QC) and average glandular dose at 79 facilities in Hokkaido in 2009, and the results of the survey were compared with those of 2004. We found that digital mammography techniques were widely used across Hokkaido and that computed radiography (CR) systems were quite widespread, with 70% of facilities having them. The average glandular dose ranged from 1.04 to 2.3 mGy (mean: 1.73 mGy) for digital equipment. The results revealed several problems. Although the use of 1-, 2-, and 3-megapixel (MP) liquid crystal displays (LCDs) was not uncommon, 5-MP LCDs were used in most cases when reading digital mammograms. Facilities that have mammography equipment are unlikely to have quality control instruments for mammography. Although daily QC is performed in most facilities, further quality control for digital mammography should be developed, including that for monitors. In a second study, we evaluated the 1 Shot Phantom M Plus (1 Shot Phantom), which enables objective evaluation by providing for one physical measurement rather than a subjective visual analysis. The results indicated that the 1 Shot Phantom was very useful for digital mammography systems in daily QC testing because it enabled objectivity.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the relation between breast glandular absorbed dose and radiographic quality in mammography

Background: Breast is one of the main radiological sensitive organs, hence it is important to evaluate absorbed dose to this organ especially to the glandular parts. It is the aim of this study to measure mean glandular dose and image contrast in terms of different mammographic parameters. Materials and Methods: In this study two mammography machines located at Said-al Shohada (Giotto) ...

متن کامل

ارزیابی دز و کیفیت تصویر در آزمون های ماموگرافی در مراکز ماموگرافی شهر تهران

Abstract Background: Breast cancer is one of the most common form of cancer among women. Diagnosis in the early stage of emergence is the most effective way to reduce the mortality caused by this disease. Mammography, the safest and most accurate imaging test for the identification of breast cancer, required high quality images along applying the lowest possible dose to sensitive tissues of the...

متن کامل

Breast dose in conventional and digital mamography

Introduction: The amount of doses received in mammography is one of the biggest concerns in the world. According to some statistics, every 12 minutes a woman dies due to breast cancer. Mammography concerned to detect breast cancer at an early stage of the disease as well as for screening, however due to the risk of exposure, the increasingly and widespread use of mammography fo...

متن کامل

نتایج یک کُد فرترن برای محاسبه دوز غده‌ای در ماموگرافی با استفاده از پارامترهای Sobol-Wu

Background: Accurate computation of the radiation dose to the breast is essential to mammography. Various the thicknesses of breast, the composition of the breast tissue and other variables affect the optimal breast dose. Furthermore, the glandular fraction, which refers to the composition of the breasts, as partitioned between radiation-sensitive glandular tissue and the adipose tissue, also h...

متن کامل

Calculation of Absorbed Glandular Dose using a FORTRAN Program Based on Monte Carlo X-ray Spectra in Mammography

Introduction: Average glandular dose calculation in mammography with Mo-Rh target-filter and dose calculation for different situations is accurate and fast. Material and Methods: In this research, first of all, x-ray spectra of a Mo target bombarded by a 28 keV electron beam with and without a Rh filter were calculated using the MCNP code. Then, we used the Sobol-Wu parameters to write a FORTRA...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Nihon Hoshasen Gijutsu Gakkai zasshi

دوره 67 4  شماره 

صفحات  -

تاریخ انتشار 2011