Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

نویسندگان

  • Guido Hilgers
  • Tonnis Nuver
  • André Minken
چکیده

In external beam radiotherapy treatment planning for patients with thoracic malignancies, respiratory-correlated CT (4D CT) is used to obtain high quality studies in the presence of respiratory motion. When helical 4D CT scans are acquired with a Brilliance CT Big Bore, the pitch must meet two conditions. It must be low enough to avoid motion artifacts, and high enough to cover the entire scan length within 120 s to prevent overheating of the X-ray tube. We developed a nomogram that can be used to obtain a suitable pitch satisfying both requirements. We also assessed the effects on the image quality of a pitch that exceeds the maximum pitch, and of a field of view (FOV) reduction. It was shown that, for AVG and MIP reconstructions, the manufacturer's maximum pitch equation yields an underestimation due to its FOV term.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment

Image-guided radiation therapy using cone-beam computed tomography (CBCT) is becoming routine practice in modern radiation therapy. The purpose of this work was to develop an imaging QA program for CT and CBCT units in our department, based on the American College of Radiology (ACR) CT accreditation phantom. The phantom has four testing modules, permitting one to test CT number accuracy, slice ...

متن کامل

Four-dimensional cardiac image by helical computed tomography.

Ultrasound and MRI are useful in assessing wall motion and systolic thickening with 2D images. Volumetric data of the helical CT provide frozen 3D images. We sought to determine whether helical CT could assess left ventricular wall motion with 3D animation. We have developed a new cardiac application, ventriculography by 4D enhanced helical CT (4D-CT-VG), which can assess wall motion with 3D an...

متن کامل

Reduction of breathing irregularity-related motion artifacts in low-pitch spiral 4D CT by optimized projection binning

BACKGROUND Respiration-correlated CT (4D CT) is the basis of radiotherapy treatment planning of thoracic and abdominal tumors. Current clinical 4D CT images suffer, however, from artifacts due to unfulfilled assumptions concerning breathing pattern regularity. We propose and evaluate modifications to existing low-pitch spiral 4D CT reconstruction protocols to counteract respective artifacts. ...

متن کامل

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Introduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, there are potential risks in addition to the clear potential benefits. Materials and Methods Two refe...

متن کامل

[Examination of motion artifacts for helical and non-helical scanning modes in head CT].

For head computed tomography (CT), non-helical scanning has been recommended even in the widely used multi-slice CT (MSCT). Also, an acute stroke imaging standardization group has recommended the non-helical mode in Japan. However, no detailed comparison has been reported for current MSCT with more than 16 slices. In this study, we compared the non-helical and helical modes for head CT, focusin...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015