Rotational Stability of Scaphoid Osteosyntheses: An In Vitro Comparison of Small Fragment Cannulated Screws to Novel Bone Screw Sets

نویسندگان

  • Jochen Erhart
  • Ewald Unger
  • Philip Schefzig
  • Peter Varga
  • Inga Trulson
  • Anna Gormasz
  • Alexander Trulson
  • Martin Reschl
  • Michael Hagmann
  • Vilmos Vecsei
  • Winfried Mayr
چکیده

BACKGROUND The current standard of care for operative repair of scaphoid fractures involves reduction and internal fixation with a single headless compression screw. However, a compression screw in isolation does not necessarily control rotational stability at a fracture or nonunion site. The single screw provides rotational control through friction and bone interdigitation from compression at the fracture site. We hypothesize that osteosyntheses with novel bone screw sets (BSS) equipped with anti-rotational elements provide improved rotational stability. METHODS Stability of osteosynthesis under increasing cyclic torsional loading was investigated on osteotomized cadaveric scaphoids. Two novel prototype BSS, oblique type (BSS-obl.) and longitudinal type (BSS-long.) were compared to three conventional screws: Acutrak2®mini, HCS®3.0 and Twinfix®. Biomechanical tests were performed on scaphoids from single donors in paired comparison and analyzed by balanced incomplete random block design. Loading was increased by 50 mNm increments with 1,000 cycles per torque level and repeated until a rotational clearance of 10°. Primary outcome measure was the number of cycles to 10° clearance, secondary outcome measure was the maximum rotational clearance for each torque level. FINDINGS BSS-obl. performed significantly better than Acutrak2®mini and HCS® (p = 0.015, p<0.0001). BSS-long. performed significantly better than HCS® (p = 0.010). No significant difference in performance between BSS-obl. and BSS-long. (p = 0.361), between BSS obl. and Twinfix® (p = 0.50) and BSS long. and Twinfix® (p = 0.667) was detected. Within the torque range up to 200 mNm, four of 21 (19%) BSS-long. and four of 21 (19%) BSS-obl. preparations showed early failure. The same loading led to early failure in four (29%) Twinfix®, seven (50%) Acutrak2®mini and 10 (71%) HCS® of 14 screw samples, respectively. CONCLUSIONS For both BSS and to a lesser extent for Twinfix® (as dual-component screw), higher rotational stabilities were identified in comparison to single component headless compression screws.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Minimal invasive treatment for scaphoid fractures using the cannulated herbert screw system.

Internal fixation of scaphoid fractures avoids the problems associated with prolonged plaster immobilization and, at the same time, allows an early return to activity for these mostly young patients.Internal fixation of the scaphoid is greatly facilitated by the use of specially designed headless bone screws, such as the Herbert bone screw, originally developed specifically for internal fixatio...

متن کامل

Scaphoid Fractures: A Comparison of Two Surgical Methods for Internal Fixation Using Either Herbert Screws or Multiple Pins

Background and Aims: Scaphoid fractures are the most common fractures of the wrist. This study compared the outcomes of two surgical internal fixation techniques, using either Herbert screws (HS group) or multiple pins (MP group), in the treatment of scaphoid fractures. Methods: This cross-sectional study was performed on 41 patients (23 in HS and 18 in MP groups) with unilateral scaphoid fract...

متن کامل

Comparison of Fixation Methods for Scaphoid Nonunions

The purpose of this study was to analyze the relative biomechanical stability of three types of internal fixation with cancellous bone graft in a cadaveric, scaphoid nonunion model. A scaphoid nonunion model was created by removing a volar wedge of bone from the waist of the scaphoid in 18 fresh frozen human cadavers. Cancellous sawbone graft was inserted into the osteotomy site and three group...

متن کامل

Impact of Screw Length on Fixed Proximal Scaphoid Fracture Biomechanics: In Vitro Study with Cyclic Loading and Load to Failure

Introduction: Internal fixation used as treatment for scaphoid fractures has increased with advances in surgical techniques [1]. Proximal pole scaphoid fractures present a clinical challenge due to their lower success rate when cast immobilization and internal fixation are utilized [2]. Screws maximizing length have been shown to impact wrist motion for fixed scaphoid waist fractures [3]. Addit...

متن کامل

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

  During fixation of a femoral neck fracture in a 23-year-old male patient with cannulated screws, a broken guide wire fragment inadvertently advanced through the hip joint and protruded into the pelvis. A second surgical approach was needed to remove the broken fragment from the pelvis. Awareness of such a potentially devastating complication will make surgeons more cautious during implementa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2016