Dose Calculation Accuracy of AAA and AcurosXB Algorithms for Small Central and Interface Lung Lesions - Verification with Gafchromic Film Dosimetry
نویسندگان
چکیده
Dose calculation for small field radiotherapy with heterogeneity often involves discrepancies, so that algorithms used by treatment planning systems (TPS) should be evaluated with reference to achieving optimal treatment results. Accuracy of two model based algorithms, AcurosXB (AcXB) and the analytical anisotropic algorithm (AAA) from Eclipse TPS, were here tested. Measurements are made using Gafchromic EBT3 films with indigenously generated lung phantoms irradiated with 6 MV photons. Lung phantoms contained two types of tumor plugs, one kept at an interface attached to the chest wall in right lung (RIT) and the other at the centre of the left lung (LCT). RIT and LCT were studied with two different tumor diameters, 1.5 cm and 2.5 cm. Scanned images were planned in TPS with 3D-CRT, IMRT and VMAT and individual plans for each tumor were irradiated keeping the Gafchromic film at the centre of the tumor to evaluate the dose distribution in the central plane. Both algorithms, irrespective of delivery techniques, showed more deviation with smaller than larger diameter tumors. Also, both demonstrated maximum deviation at the junction of tumor and lung in both RIT and LCT cases. However, the deviation observed was higher with AAA and a minimal acceptable deviation of within 4 % was achieved with AcurosXB.
منابع مشابه
Evaluation of AAA and XVMC Algorithms for Dose Calculation in Lung Equivalent Heterogeneity in Photon Fields: A Comparison of Calculated Results with Measurements
Aims: The aims of the present work are (1) to evaluate dose calculation accuracy of two commonly used algorithms for 15 MV small photon fields in a medium encompassing heterogeneity and (2) to compare them with measured results obtained from gafchromic film EBT2.Materials and Methods: Authors employed kailwood (Pinus Wallichiana) to mimic lung. Briefly, seven Kailwood plates, each measuring 25x...
متن کاملdimensional film dosimetry with GAFCHROMIC films for quality assurance and dosimetric verification of 3D conformal radiotherapy in the presence of heterogeneities
Introduction: The presence of heterogeneities such as air-filled cavities in the head and neck treatment fields region, may result in potential dosimetric disagreement because the losses of charged particle equilibrium. Most of treatments planning systems are not able to predict dose distribution of inhomogeneities region accurately. Therefore, dose calculation algorithms need to...
متن کاملEvaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry
Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...
متن کاملSmall photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملDosimetric accuracy of the Acuros XB and Anisotropic analytical algorithm near interface of the different density media for the small fields of a 6- MV flattening-filter-free beam
Background: This study was conducted to assess the accuracy of dose calculation near the air-phantom interface of a heterogeneous phantom for Acuros XB (AXB) and Anisotropic Analytical Algorithm (AAA) algorithm of a 6-MV flattening-filter-free beam, compared with film measurements. Materials and Methods: A phantom included air gap was ...