Fluid dynamic analysis of the 50 cc Penn State artificial heart under physiological operating conditions using particle image velocimetry.

نویسندگان

  • Pramote Hochareon
  • Keefe B Manning
  • Arnold A Fontaine
  • John M Tarbell
  • Steven Deutsch
چکیده

In order to bridge the gap of existing artificial heart technology to the diverse needs of the patient population, we have been investigating the viability of a scaled-down design of the current 70 cc Penn State artificial heart. The issues of clot formation and hemolysis may become magnified within a 50 cc chamber compared to the existing 70 cc one. Particle image velocimetry (PIV) was employed to map the entire 50 cc Penn State artificial heart chamber. Flow fields constructed from PIV data indicate a rotational flow pattern that provides washout during diastole. In addition, shear rate maps were constructed for the inner walls of the heart chamber The lateral walls of the mitral and aortic ports experience high shear rates while the upper and bottom walls undergo low shear rates, with sufficiently long exposure times to potentially induce platelet activation or thrombus formation. In this study, we have demonstrated that PIV may adequately map the flow fields accurately in a reasonable amount of time. Therefore, the potential exists of employing PIV as a design tool.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Wall Shear - Rate Estimation Within the 50 cc Penn State Artificial Heart Using Particle Image

Particle image velocimetry (PIV) has been gaining acceptance as a routine tool to evaluate the flow fields associated with fluid mechanical devices. We have developed algorithms to investigate the wall shear-rates within the 50cc Penn State artificial heart using low magnification, conventional particle image velocimetry (PIV). Wall shear has been implicated in clot formation, a major post-impl...

متن کامل

Fluid Dynamics Investigation of a GDI Fuel Spray by Particle Image Velocimetry

In this work, result of experimental investigation on interaction of fuel spray generated by a swirled type injector, with air motion in a prototype cylinder are presented. Experiments were carried out by planar imaging and particle image velocimetry (PIV) techniques in order to provide information about the spray structure evolution and instantaneous velocity distribution of air motion and ...

متن کامل

Correlation of in vivo clot deposition with the flow characteristics in the 50 cc penn state artificial heart: a preliminary study.

Flow stasis in an artificial heart may provide a situation where thrombus develops. Should part, or all, of the clot dislodge, a thromboembolism may lead to stroke(s), neurologic deficits, or even death. In an effort to determine if the regime of low shear or stasis exists, a two-dimensional particle image velocimetry (PIV) system was implemented to measure the velocity field within the 50 cc P...

متن کامل

Off-design considerations of the 50cc Penn State Ventricular Assist Device.

Particle image velocimetry is used to study the flow field associated with off-design conditions of the 50cc Penn State Ventricular Assist Device. Phase average data, for 300 consecutive beats, was acquired for 60, 75, and 90 beats per minute (bpm) at 35/65 systolic duration and at 75 bpm for 50/50 systolic duration. Diastolic flow is dominated by the inflow jet and shows more penetration and b...

متن کامل

Determination of Suitable Operating Conditions of Fluid Catalytic Cracking Process by Application of Artificial Neural Network and Firefly Algorithm

Fluid Catalytic Cracking (FCC) process is a vital unit to produce gasoline. In this research, a feed forward ANN model was developed and trained with industrial data to investigate the effect of operating variables containing reactor temperature feed flow rate, the temperature of the top of the main column and the temperature of the bottom of the debutanizer tower on quality and quantity of...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomechanical engineering

دوره 126 5  شماره 

صفحات  -