In vitro evaluation of heat and moisture exchangers designed for spontaneously breathing tracheostomized patients.

نویسندگان

  • Claudia Brusasco
  • Francesco Corradi
  • Maria Vargas
  • Margherita Bona
  • Federica Bruno
  • Maria Marsili
  • Francesca Simonassi
  • Gregorio Santori
  • Paolo Severgnini
  • Robert M Kacmarek
  • Paolo Pelosi
چکیده

BACKGROUND Heat and moisture exchangers (HMEs) are commonly used in chronically tracheostomized spontaneously breathing patients, to condition inhaled air, maintain lower airway function, and minimize the viscosity of secretions. Supplemental oxygen (O2) can be added to most HMEs designed for spontaneously breathing tracheostomized patients. We tested the efficiency of 7 HMEs designed for spontaneously breathing tracheostomized patients, in a normothermic model, at different minute ventilations (VE) and supplemental O2 flows. METHODS HME efficiency was evaluated using an in vitro lung model at 2 VE (5 and 15 L/min) and 4 supplemental O2 flows (0, 3, 6, and 12 L/min). Wet and dry temperatures of the inspiratory flow were measured, and absolute humidity was calculated. In addition, HME efficiency at 0, 12, and 24 h use was evaluated, as well as resistance to flow at 0 and 24 h. RESULTS The progressive increase in O2 flow from 0 to 12 L/min was associated with a reduction in temperature and absolute humidity. Under the same conditions, this effect was greater at lower VE. The HME with the best performance provided an absolute humidity of 26 mg H2O/L and a temperature of 27.8 °C. No significant changes in efficiency or resistance were detected during the 24 h evaluation. CONCLUSIONS The efficiency of HMEs in terms of temperature and absolute humidity is significantly affected by O2 supplementation and V(E).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Humidification performance of humidifying devices for tracheostomized patients with spontaneous breathing: a bench study.

BACKGROUND Heat and moisture exchangers (HMEs) are commonly used for humidifying respiratory gases administered to mechanically ventilated patients. While they are also applied to tracheostomized patients with spontaneous breathing, their performance in this role has not yet been clarified. We carried out a bench study to investigate the effects of spontaneous breathing parameters and oxygen fl...

متن کامل

Breathing-swallowing interaction in neuromuscular patients: a physiological evaluation.

RATIONALE Malnutrition and aspiration are major problems in patients with neuromuscular disease. Because impaired swallowing contributes to malnutrition, means of improving swallowing are needed. OBJECTIVES To investigate interactions between breathing and swallowing in neuromuscular disorders and to evaluate the impact of mechanical ventilation (MV) on swallowing in tracheostomized patients....

متن کامل

Prevention of Facial Cold Injury with

The present study evaluated the thermodynamic characteristics of a prototype respiratory heat and moisture exchanger (HMS), particularly its ability to prevent cold injury to the face. The HME comprised a thermally insulated oro-nasal mask with a cylindrical heat and moisture exchanger protruding from the centre of the mask. The heat exchanging unit contained an aluminium “honeycomb” structure ...

متن کامل

Thermal Design Considerations and Performance Evaluation of Cryogenic Tube in Tube Heat Exchangers

Heat exchangers are the most important equipment in refrigeration processes. Design and modeling of heat exchangers operating at low temperatures are different from other regular heat exchangers. This study includes two sections. In the first section, design and modeling considerations needed for evaluating the real thermal behavior of heat exchangers at low temperatures were discussed. The...

متن کامل

Effectiveness of a heat and moisture exchanger in preventing hyperpnoea induced bronchoconstriction in subjects with asthma.

The effect of a heat and moisture exchanger, a device with hygroscopic material for conditioning inspired air, on hyperpnoea induced bronchoconstriction was studied in nine non-smoking volunteers with asthma, aged 19-32 years. Each had previously shown an increase of at least 100% in specific airways resistance (sRaw) to isocapnic hyperpnoea with dry air. On two separate days the subject perfor...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 58 11  شماره 

صفحات  -

تاریخ انتشار 2013