Scatter and Cross-Talk Corrections in Simultaneous Tc-99m/I-123 Brain SPECT using Constrained Factor Analysis and Artificial Neural Networks1

نویسندگان

  • G. El Fakhri
  • P. Maksud
  • M. F. Kijewski
  • M. O. Habert
  • A. Todd-Pokropek
  • S. C. Moore
چکیده

Simultaneous imaging of Tc-99m and I-123 would have a high clinical potential in the assessment of brain perfusion (Tc-99m) and neurotransmission (I-123) but is hindered by cross-talk between the two radionuclides. Monte Carlo simulations of 15 different dual-isotope studies were performed using a digital brain phantom. Several physiologic Tc-99m and I-123 uptake patterns were modeled in the brain structures. Two methods were considered to correct for cross-talk from both scattered and unscattered photons: constrained spectral factor analysis (SFA) and artificial neural networks (ANN). The accuracy and precision of reconstructed pixel values within several brain structures were compared to those obtained with an energy windowing method (WSA). In I-123 images, mean bias was close to 10% in all structures for SFA and ANN and between 14% (in the caudate nucleus) and 25% (in the cerebellum) for WSA. Tc-99m activity was overestimated by 35% in the cortex and 53% in the caudate nucleus with WSA, but by less than 9% in all structures with SFA and ANN. SFA and ANN performed well even in the presence of highenergy I-123 photons. The accuracy was greatly improved by incorporating the contamination into the SFA model or in the learning phase for ANN. SFA and ANN are promising approaches to correct for cross-talk in simultaneous Tc-99m/I123 SPECT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Simultaneous Tc-99m and I-123 dual-radionuclide imaging with a solid-state detector-based brain-SPECT system and energy-based scatter correction

BACKGROUND A brain single-photon emission computed tomography (SPECT) system using cadmium telluride (CdTe) solid-state detectors was previously developed. This CdTe-SPECT system is suitable for simultaneous dual-radionuclide imaging due to its fine energy resolution (6.6 %). However, the problems of down-scatter and low-energy tail due to the spectral characteristics of a pixelated solid-state...

متن کامل

Simultaneous technetium-99m/thallium-201 SPECT imaging with model-based compensation for cross-contaminating effects.

Simultaneous acquisition of dual-isotope SPECT data offers a number of advantages over separately acquired data; however, simultaneous acquisition can result in cross-contamination between isotopes. In this work we propose and evaluate two frameworks for iterative model-based compensation of cross-contamination in dual-isotope SPECT. The methods were applied to cardiac imaging with technetium-9...

متن کامل

Dual radioisotopes simultaneous SPECT of 99mTc-tetrofosmin and 123I-BMIPP using a semiconductor detector

OBJECTIVES The energy resolution of a cadmium-zinc-telluride (CZT) solid-state semiconductor detector is about 5%, and is superior to the resolution of the conventional Anger type detector which is 10%. Also, the window width of the high-energy part and of the low-energy part of a photo peak window can be changed separately. In this study, we used a semiconductor detector and examined the effec...

متن کامل

Accuracy of quantitative reconstructions in SPECT/CT imaging.

The goal of this study was to determine the quantitative accuracy of our OSEM-APDI reconstruction method based on SPECT/CT imaging for Tc-99m, In-111, I-123, and I-131 isotopes. Phantom studies were performed on a SPECT/low-dose multislice CT system (Infinia-Hawkeye-4 slice, GE Healthcare) using clinical acquisition protocols. Two radioactive sources were centrally and peripherally placed insid...

متن کامل

Spectral Factor Analysis for Multi-isotope Imaging in Nuclear Medicine

In nuclear medicine, simultaneous dual-isotope imaging is used to determine the distribution of two radiotracers from a single acquisition and for emission/transmission (E/T) imaging in SPECT. However, no general solution to the cross–talk problem caused by scattered and unscattered photons has been found yet and accurate quantification cannot be performed. We describe a general method of spect...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009