Changes of nebulizer output over the years.

نویسندگان

  • P J Merkus
  • E E van Essen-Zandvliet
  • E Parlevliet
  • G Borsboom
  • P J Sterk
  • K F Kerrebijn
  • P H Quanjer
چکیده

The effects of long-term use and of cleaning on the output of 30 specimens of DeVilbiss 646 nebulizers were investigated in order to assess their influence on outcome and interpretation of inhalation provocation tests. Output was assessed in a standardized manner based on weight loss on four occasions one year apart. Output was highly reproducible (intraclass correlation greater than or equal to 0.90), and varied considerably between nebulizers, necessitating calibration at least once. Inadequate cleaning diminished nebulizer output by a factor of two (p less than 0.01), while long-term use was associated with a moderate increase in output of 0.28 mg.yr-1 (p less than 0.0001). Bronchial responsiveness can be underestimated by about one doubling dose due to inadequate cleaning, while interchanging nebulizers can lead to overestimation or underestimation by up to one doubling dose. With proper care the increase in output due to wear has no consequences for clinical practice, or for longitudinal studies.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of tidal volume on the output and particle size distribution of hypertonic saline from an ultrasonic nebulizer.

The purpose of the present study was to measure the output and particle size distribution of hypertonic saline from an ultrasonic nebulizer over the range of tidal volumes (VT) used in an epidemiological survey of asthma in children. A Starling pump was used to generate different VTs from 250 to 500 ml, at different respiratory rates of 12 and 20 breaths.min-1, to cover the normal range of resp...

متن کامل

Comparison of twenty three nebulizer/compressor combinations for domiciliary use.

We have assessed the physical and dynamic characteristics of 23 home jet nebulizer/compressor combinations currently available in the UK and Europe. The combinations were evaluated in terms of pressure-flow characteristics, aerosol mass distribution, volume output, electrical costs, and sound level. In addition, we determined the effect of nebulizer fill volume on aerosol mass distribution and ...

متن کامل

Comparison of the effect of broncodiatory albotrol from spacer and nebulazor foe freatment of Astma patients

Tavana S1, Nadi E1, Ansari M2, zeraati F3, Hashemi H1, Mani Kashani Kh4, Mavallat M5 1. Assistant professor, Department of internal, Faculty of medicine, Hamedan University of medical sciences 2. Associate professor, Department of internal, Faculty of medicine, Hamedan University of medical sciences 3. Associate professor, Department of Pharmacology, Faculty of medicine, Hamedan Univers...

متن کامل

Comparison of a Nebulizer Using a Novel Aerosol Generator with a Standard Ultrasonic Nebulizer Designed for Use during Mechanical Ventilation

Figure 1: Diagram of SUN 145 (from product instruction manual) Ultrasonic nebulizers (USNs) generate aerosols without changing ventilator volumes or alarms. However, USN output varies with gas flow, is position dependent and generates heat. We wanted to determine whether the AeronebTM Professional Nebulizer System (Aeroneb Pro; Aerogen, Inc., Sunnyvale, CA)* has similar limitations. METHODS: We...

متن کامل

Output characteristics of DeVilbiss No. 40 hand-held jet nebulizers.

DeVilbiss No. 40 hand-held nebulizers are widely used for quantifying airway responsiveness in large populations using pharmacological agents. We examined the aerosol characteristics of five nebulizers. Within each device, the aerosol output and droplet size were reasonably stable over a wide range of bulb pressures, although there were considerable differences in output characteristics between...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 5 4  شماره 

صفحات  -

تاریخ انتشار 1992