A comparison between two electronic apex locators: an ex vivo investigation.
نویسندگان
چکیده
AIM To compare ex vivo the performance of the Apex Finder and the Root ZX apex locators, with and without irrigant, in canals having different diameters. METHODOLOGY Sixty canals in 60 teeth were prepared using stainless steel hand files and 0.04 taper NiTi rotary instruments. During preparation the narrowest diameter of the canal was transported to the apical root surface. The canals were irrigated with RC-Prep and 5% NaOCl solution. Six groups were obtained, each with 10 canals having the same diameter of foramen, either 0.15, 0.20, 0.25, 0.40, 0.60 and 0.80 mm. A size 15 K-file was advanced into each canal until its tip was observed under x10 magnification to reach the foramen and the corresponding length recorded. The measurements were performed to an accuracy of 0.25 mm as a base unit of length. The teeth were then fixed to a plastic bar suspended over a glass container filled with 0.9% NaCl solution. Each apex locator was tested when the K-file was at the foramen, or 0.5, 1.0, 1.5 and 2.0 mm short; with the root apex immersed into the solution; with the canal dry or irrigated with NaCl. To evaluate the accuracy of both electronic apex locators (EALs) each electronically determined distance was compared with the actual length and the data analysed using the General Linear Model and the Student t-test. RESULTS Out of 2400 measurements 100 were electrically unstable, all with the Root ZX. In total, 521 measurements located the position of the file tip beyond the apex, in general, in high conductive conditions with the Root ZX and in low conductive conditions with the Apex Finder. No significant difference in terms of accuracy was found between the two EALs when the file tip was at the foramen (Root ZX mean +0.12 mm, SD 1.22 mm; Apex Finder mean +0.57 mm, SD 1.16 mm). Comparing all the measurements performed with the file tip within 2 mm of the foramen, in all the different conditions tested, the accuracy was affected (P<0.025) by diameter of the foramen, type of EAL, distance to the apex, and by several interactions. CONCLUSIONS Under the different ex vivo conditions both EALs provided accurate measurements when the file tip was at the foramen. The accuracy of the Apex Finder was negatively influenced by high conductive conditions, whilst the Root ZX provided inaccurate and unstable measurements mostly in low conductive conditions.
منابع مشابه
Evaluation of the accuracy of two apex locators in endodontic treatment and retreatments: an ex vivo study
Introduction: Working length determination is important in successful endodontic treatment and retreatment.This study evaluated the accuracy of two electronic apex locators Root ZX and Raypex®6 (EALs) in determining the electronic working length (EWL) of the root canals in endodontic treatment and retreatment. Materials and Methods: Access cavities were prepared on forty extracted, singl...
متن کاملIn vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملEx vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.
AIM To evaluate ex vivo the accuracy of two electronic apex locators during root canal length determination in primary incisor and molar teeth with different stages of physiological root resorption. METHODOLOGY One calibrated examiner determined the root canal length in 17 primary incisors and 16 primary molars (total of 57 root canals) with different stages of root resorption based on the ac...
متن کاملThe effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants
BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...
متن کاملمقایسه دقت دستگاههای اپکس یاب الکترونیکی ZX Root،Novapex در مقایسه با رادیوگرافی در شرایط in vivo
Background and Aims: Working length determination and remaining this length is of great importance in root canal therapy. Recently, electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the sensitivity of Root ZX and Novapex electronic apex locators with radiography in a clinical study. Materials and...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- International endodontic journal
دوره 40 5 شماره
صفحات -
تاریخ انتشار 2007