Cervical total disc replacement using a novel compressible prosthesis: Results from a prospective Food and Drug Administration–regulated feasibility study with 24-month follow-up

نویسندگان

  • Carl Lauryssen
  • Domagoj Coric
  • Thomas Dimmig
  • David Musante
  • Donna D. Ohnmeiss
  • Harrison A. Stubbs
چکیده

BACKGROUND Anterior cervical fusion, an established procedure to treat cervical radiculopathy, sacrifices the natural function of the disc, while placing increased stresses on adjacent spinal levels. In contrast, the cervical total disc replacement (cTDR) maintains motion and decreases adjacent-level stresses. The purpose of this study was to investigate the safety and effectiveness of a next-generation cTDR device in patients with symptomatic cervical radiculopathy. METHODS This is a multicenter Food and Drug Administration-regulated feasibility study to evaluate safety and effectiveness of the M6-C Artificial Cervical Disc for the treatment of patients with symptomatic cervical radiculopathy at 1 or 2 levels from C3 to C7. Neck Disability Index (NDI), visual analog scales (VAS) assessing neck and arm pain, Short Form 36 Health Survey (SF-36), safety, and radiographic outcomes were assessed preoperatively, at 6 weeks and 3, 6, 12, and 24 months postoperatively. RESULTS Thirty patients were enrolled at 3 clinical sites. Patients were implanted at either 1 or 2 levels. Mean NDI improved from 67.8 to 20.8 (P < .0001) at 24 months. Significant improvement was also observed through 24-month follow-up in neck and arm pain VAS (P < .0001) and in physical (P < .005) and mental component scores of the SF-36 at 3, 6, and 12 months (P < .008). There were no serious adverse events related to the device or procedure as adjudicated by an independent clinical events committee. Radiographically, disc space height increased more than 50% with a correlative increase in the postoperative disc angle. Range of motion decreased slightly from baseline during early follow-up but increased slightly and were maintained throughout the follow-up period. CONCLUSIONS The M6-C cervical artificial disc represents a new generation of cTDR design. Results of this study found the M6-C device to produce positive clinical and radiographic outcomes similar to other cTDRs, warranting further investigation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Two- to 3-Year Follow-Up of ProDisc-L: Results From a Prospective Randomized Trial of Arthroplasty Versus Fusion

BACKGROUND Although total disc replacement has been performed for years outside the United States, relatively little available data address clinical outcomes, particularly data from prospective studies. We report the 24- to 36-month follow-up of one center's experience with the ProDisc-L artificial disc as part of a prospective, randomized trial comparing total disc arthroplasty to combined ant...

متن کامل

Multi-center, Prospective, Randomized, Controlled Investigational Device Exemption Clinical Trial Comparing Mobi-C Cervical Artificial Disc to Anterior Discectomy and Fusion in the Treatment of Symptomatic Degenerative Disc Disease in the Cervical Spine

BACKGROUND Anterior cervical discectomy and fusion (ACDF) is the gold standard for treating symptomatic cervical disc degeneration. Cervical total disc replacements (TDRs) have emerged as an alternative for some patients. The purpose of this study was to evaluate the safety and effectiveness of a new TDR device compared with ACDF for treating single-level cervical disc degeneration. METHODS T...

متن کامل

The Mobi-C cervical disc for one-level and two-level cervical disc replacement: a review of the literature

BACKGROUND Cervical disc arthroplasty (CDA) is a novel motion-preserving procedure that is an alternative to fusion. The Mobi-C disc prosthesis, one of many Food and Drug Administration (FDA)-approved devices for CDA, is the only FDA-approved prosthesis for two-level CDA. Hence, it may allow for improved outcomes compared with multilevel fusion procedures. PURPOSE To critically assess the ava...

متن کامل

Kineflex lumbar artificial disc versus Charité lumbar total disc replacement for the treatment of degenerative disc disease: A randomized non-inferiority trial with minimum of 2 years’ follow-up

BACKGROUND The Kineflex lumbar artificial disc replacement device (SpinalMotion, Mountain View, California) is a semiconstrained, posterior center of rotation, metal-on-metal intervertebral disc prosthesis. We performed a prospective, randomized, non-inferiority trial comparing the Kineflex Disc with the Food and Drug Administration (FDA)-approved Charité device (DePuy Spine, Raynham, Massachus...

متن کامل

Comparison of Titanium vs. Polycel Total Ossicular Replacement Prosthesis

Introduction: Even though modern technology progresses so rapidly, annals of otology are replete with so many challenging article, which often compare various types of prosthesis. Since there has not been a prospective randomized clinical trial study which compares the hearing result of total ossicular replacement prosthesis made of Titanium with omega connector and Polycel in the literature, w...