Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM): An Ovine Cardiac Output Validation

نویسندگان

  • Robert A. Phillips
  • Sally G. Hood
  • Beverley M. Jacobson
  • Malcolm J. West
  • Li Wan
  • Clive N. May
چکیده

Background. The pulmonary artery catheter (PAC) is an accepted clinical method of measuring cardiac output (CO) despite no prior validation. The ultrasonic cardiac output monitor (USCOM) is a noninvasive alternative to PAC using Doppler ultrasound (CW). We compared PAC and USCOM CO measurements against a gold standard, the aortic flow probe (FP), in sheep at varying outputs. Methods. Ten conscious sheep, with implanted FPs, had measurements of CO by FP, USCOM, and PAC, at rest and during intervention with inotropes and vasopressors. Results. CO measurements by FP, PAC, and USCOM were 4.0 ± 1.2 L/min, 4.8 ± 1.5 L/min, and 4.0 ± 1.4 L/min, respectively, (n = 280, range 1.9 L/min to 11.7 L/min). Percentage bias and precision between FP and PAC, and FP and USCOM was -17 and 47%, and 1 and 36%, respectively. PAC under-measured Dobutamine-induced CO changes by 20% (relative 66%) compared with FP, while USCOM measures varied from FP by 3% (relative 10%). PAC reliably detected -30% but not +40% CO changes, as measured by receiver operating characteristic area under the curve (AUC), while USCOM reliably detected ±5% changes in CO (AUC > 0.70). Conclusions. PAC demonstrated poor accuracy and sensitivity as a measure of CO. USCOM provided equivalent measurements to FP across a sixfold range of outputs, reliably detecting ±5% changes.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the USCOM ultrasound cardiac output monitor with pulmonary artery catheter thermodilution in patients undergoing liver transplantation.

The aim of the study was to compare the standard technique of cardiac output determination by pulmonary artery catheter thermodilution (PAC-TD) with a noninvasive ultrasound Doppler monitor (USCOM Pty., Ltd., Coffs Harbour, Australia) in surgery for liver transplantation. We wished to determine if the degree of accuracy would allow the ultrasound cardiac output monitor (USCOM) to be used as an ...

متن کامل

Comparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter.

BACKGROUND Cardiac output (CO) cannot reliably be estimated by clinical examination. We aimed to measure the agreement between CO measurements using a supra-sternal Doppler monitor (USCOM, Coffs Harbour, Australia) and the pulmonary artery catheter (PAC). METHODS The study was conducted in the intensive care unit of a tertiary teaching hospital. All patients with PAC in situ were eligible. Si...

متن کامل

Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit.

BACKGROUND The USCOM ultrasonic cardiac output monitor (USCOM Pty Ltd, Coffs Harbour, NSW, Australia) is a non-invasive device that determines cardiac output by continuous-wave Doppler ultrasound. The aim of this study was to evaluate the accuracy of the USCOM device compared with the thermodilution technique in intensive care patients who had just undergone cardiac surgery. METHODS We conduc...

متن کامل

Pulmonary artery catheterization and esophageal doppler monitoring in the ICU.

The clinical assessment of cardiac performance and ventricular preload is notoriously unreliable in critically ill patients. Consequently, a number of technologies have been developed to provide the clinician with indexes of cardiovascular function to assist in therapeutic decision making. Foremost among these is the pulmonary artery catheter (PAC). Indeed, the PAC has largely shaped the practi...

متن کامل

Comparison of continuous-wave Doppler ultrasound monitor and echocardiography to assess cardiac output in intensive care patients.

BACKGROUND Continuous-wave Doppler (CWD) ultrasound through the left ventricular outflow tract is one modality used for non-invasive cardiac output measurement. The ultrasonic cardiac output monitor (USCOM) is a relatively new monitor which uses a small, transcutaneous ultrasound probe to measure cardiac output with CWD via the suprasternal window. It is faster and less complex to train new use...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 2012  شماره 

صفحات  -

تاریخ انتشار 2012