Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems
نویسندگان
چکیده
PURPOSE This study evaluated the influence of abutment materials on the stability of the implant-abutment joint in internal conical connection type implant systems. MATERIALS AND METHODS Internal conical connection type implants, cement-retained abutments, and tungsten carbide-coated abutment screws were used. The abutments were fabricated with commercially pure grade 3 titanium (group T3), commercially pure grade 4 titanium (group T4), or Ti-6Al-4V (group TA) (n=5, each). In order to assess the amount of settlement after abutment fixation, a 30-Ncm tightening torque was applied, then the change in length before and after tightening the abutment screw was measured, and the preload exerted was recorded. The compressive bending strength was measured under the ISO14801 conditions. In order to determine whether there were significant changes in settlement, preload, and compressive bending strength before and after abutment fixation depending on abutment materials, one-way ANOVA and Tukey's HSD post-hoc test was performed. RESULTS Group TA exhibited the smallest mean change in the combined length of the implant and abutment before and after fixation, and no difference was observed between groups T3 and T4 (P>.05). Group TA exhibited the highest preload and compressive bending strength values, followed by T4, then T3 (P<.001). CONCLUSION The abutment material can influence the stability of the interface in internal conical connection type implant systems. The strength of the abutment material was inversely correlated with settlement, and positively correlated with compressive bending strength. Preload was inversely proportional to the frictional coefficient of the abutment material.
منابع مشابه
Effect of Abutment Connection Type and Cyclic Loading on Removal Torque Value
Background and Aim: Maintaining the primary torque of the abutment screw is a common problem related to implant-supported restorations; a failure in this respect often leads to screw loosening. The abutment screw plays a critical role in the long-term stability of restorations. This study aimed to assess the effect of two abutment connection types with and without cyclic loading on removal torq...
متن کاملAbutment fracture and the technique Used to Remove the fracture portion of the Abutmet: Case Report
Abstract This clinical report presents a case of abutment fracture, due to occlusal overload, occurred with an implant-supported restoration on right mandibular first molar of a female patient. After 6 months of prosthetic reconstruction implants, the patient returned to the clinic due to the loss of implant crown. The fractured fragment of the abutment with the screw protec...
متن کاملInfluence of the implant-abutment connection design and diameter on the screw joint stability
PURPOSE This study was conducted to evaluate the influence of the implant-abutment connection design and diameter on the screw joint stability. MATERIALS AND METHODS Regular and wide-diameter implant systems with three different joint connection designs: an external butt joint, a one-stage internal cone, and a two-stage internal cone were divided into seven groups (n=5, in each group). The in...
متن کاملمقایسه تشخیص رادیوگرافیک عدم تطابق اباتمنتهای زیرکونیایی و تیتانیومی در سه نوع مختلف ناحیه اتصال
Background and Aims: In response to esthetic demand and use of zirconia abutments; detection of implant-abutment connection misfit is so important. The purpose of this study was to evaluate the sensitivity and specificity of radiographic images in the detection of abutment-implant connection misfit in zirconia and titanium abutments of three different implant connections. Materials and Met...
متن کاملEvaluation of the Effect of Zirconia and Titanium Abutments on Microleakage of Implant-Abutment Interface Under Oblique Cyclic Loading In Vitro
Background and Aim: Oral bacteria can proliferate in the implant-abutment interface (IAI) and cause inflammation in the peri-implant tissues and adjacent bone. This study aimed to assess the effect of zirconia and titanium abutments on the microleakage of the IAI under oblique cyclic loading conditions. Materials and Methods: In this in-vitro study, 12 implant-abutment assemblies with zirconia...
متن کامل