Corneal modulus and IOP measurements in canine eyes using Goldmann applanation tonometry and Tono-pen.

نویسندگان

  • Junhua Tang
  • Xueliang Pan
  • Paul A Weber
  • Jun Liu
چکیده

PURPOSE To experimentally examine the effect of corneal modulus on Goldmann applanation tonometry (GAT) and Tono-pen (Tono-pen XL, Reichert, Inc., Depew, NY) measurements of intraocular pressure (IOP) in a canine eye model. METHODS Twenty-one canine globes were recovered from healthy animals. IOP was controlled at 10, 15, 20, 30, and 40 mm Hg and measured by GAT and Tono-pen following standard protocols. The corneas were dissected and uniaxial tensile tests were performed on corneal strips. The correlation between GAT and Tono-pen errors and corneal secant modulus was evaluated using Pearson correlation coefficients. The influence of corneal thickness and the true pressure was also examined. RESULTS At a true IOP of 10, 15, 20, 30, and 40 mm Hg, the GAT readings were 1.1 ± 1.0, 5.1 ± 1.5, 9.5 ± 2.0, 17.3 ± 1.6, and 25.3 ± 1.8 mm Hg, respectively. The corresponding Tono-pen readings were 7.8 ± 1.7, 12.4 ± 1.7, 16.1 ± 1.9, 22.5 ± 2.1, and 28.1 ± 2.2 mm Hg, respectively. The mean secant modulus at 1% strain of the canine corneal strips was 1.54 ± 0.43 megapascal (MPa). Corneal secant modulus was significantly correlated with GAT errors when the true IOP was 30 mm Hg (R = 0.49; P < 0.05). No significant correlation was observed between tonometric errors and corneal thickness. Both GAT and Tono-pen errors increased significantly at higher pressures (P < 0.001). CONCLUSIONS Both GAT and Tono-pen underestimated IOP in canine eyes. There was preliminary experimental evidence for a correlation between corneal modulus and GAT in the canine eyes and a higher corneal modulus was associated with higher GAT readings at a certain pressure level. The tonometric errors appeared to be pressure-dependent.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea.

AIMS To evaluate the intraocular pressure (IOP) measurements in patients with band keratopathy or glued corneas obtained from affected and non-affected areas. METHODS 15 patients with band keratopathy or glued corneas were prospectively recruited. When both eyes were affected, only the right eye was analysed. Tono-Pen readings of IOP were obtained sequentially from the affected and non-affect...

متن کامل

Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies

Unspecified Posted at the Zurich Open Repository and Archive, University of Zurich ZORA URL: http://doi.org/10.5167/uzh-101903 Published Version Originally published at: Neuburger, M; Grosswendt, J; Lautebach, S; Maier, P; Birnbaum, F; Böhringer, D; Funk, J; Jordan, J F; Reinhard, T (2014). Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral...

متن کامل

Tono-Pen tonometry in normal and in post-keratoplasty eyes.

Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 82 eyes of 82 patients with normal corneas and in 54 eyes of 54 patients who had undergone penetrating keratoplasty and whose corneas did not preclude the use of Goldmann tonometer. We found that the intraocular pressure (IOP) in 48% of the eyes with normal corneas and in 57% after keratoplasty has different measureme...

متن کامل

The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

AIMS To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT). METHODS CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending...

متن کامل

Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.

A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Investigative ophthalmology & visual science

دوره 52 11  شماره 

صفحات  -

تاریخ انتشار 2011