The Amplatzer duct occluder: experience in 209 patients.
نویسندگان
چکیده
OBJECTIVES The aim of the study was to assess the safety and efficacy of the Amplatzer ductal occluder (ADO) in transcatheter occlusion of patent ductus arteriosus (PDA). BACKGROUND Transcatheter closure of small to moderate sized PDAs is an established procedure. The ADO is a self-expandable device with a number of salutary features, notably its retrievability, ease of delivery via small 5F to 7F catheters and a range of sizes suitable even for the larger PDAs. METHODS Between November 1997 and August 1999, the ADO was successfully implanted in 205 of 209 patients with PDA. The inclusion criteria for this device occlusion method were patients with clinical and echocardiographic features of moderate to large PDA, weighing > or =3.5 kg as well as asymptomatic adolescents and adults with PDA measuring > or =5.0 mm on two-dimensional (2D) echocardiogram. Occlusion was achieved via the antegrade venous approach. Follow-up evaluations were performed with 2D echocardiogram, color-flow mapping and Doppler measurement of the descending aorta and left pulmonary artery velocity at 24 h and 1, 3, 6 and 12 months after implantation. RESULTS Two hundred and five patients had successful PDA occlusion using this device. The patients were between two months and 50 years (median 1.9) and weighed between 3.4 kg and 63.2 (median 8.4). Infants made up 26% of the total patients. The PDA measured from 1.8 to 12.5 mm (mean 4.9) at the narrowest diameter. Forty-four percent of patients achieved immediate complete occlusion. On color Doppler the closure rates at 24 h and 1 month after implant were 66% and 97%, respectively. At 6 and 12 months all except one patient attained complete occlusion. Device embolization occurred in three patients; in two this was spontaneous, and in the other it was due to catheter manipulation during postimplant hemodynamic measurement. Mild aortic narrowing was seen in an infant. CONCLUSIONS Patent ductus arteriosus occlusion using ADO is safe and efficacious. It is particularly useful in symptomatic infants and small children with relatively large PDA. Embolization can be minimized by selection of appropriate sized devices, and caution should be exercised in infants <5 kg.
منابع مشابه
Transcatheter closure of the patent ductus arteriosus using an Amplatzer duct occluder in adults.
Patent ductus arteriosus (PDA) is a congenital heart disease that frequently escapes detection until the patient reaches adulthood. Percutaneous closure of the PDA has been established as a safe alternative to surgery, the Amplatzer duct occluder (ADO) is a new device that has been recently reported in paediatric patients. Our initial experience with an of ADO in an adult population appears to ...
متن کاملTranscatheter closure of aortopulmonary window with Amplatzer duct occluder II
Sir, An aortopulmonary window (APW) is a rare cardiac malformation, first reported by Elliotson, is found in 0.1% of patients with congenital cardiac lesions.[1] In most cases, surgical repair is undertaken during infancy although isolated cases have been reported of transcatheter device closures. We report our experience of transcatheter APW closure via antegrade and retrograde methods using a...
متن کاملComplementary Techniques of Percutaneous Closure of Ductus Arteriosus Using Detachable Cook Coils and Amplatzer Devices.
BACKGROUND Patent Ductus Arteriosus (PDA) is a commoncardiac malformation whose treatment locally has been surgical ligation via a lateral thoracotomy. Device closure of the ductus was first performed at the Mater hospital in 1999 in a ten year old male using a five millilitre detachable cook coil. In 2000 the Amplatzer device was introduced to close larger ducts. Subsequently these devices hav...
متن کاملSafety and efficacy of Amplatzer duct occluder for percutaneous closure of ventricular septal defects with tunnel shape aneurysm: Medium term follow up
Objectives: Different devices including Amplatzer duct occluder has been used for percutaneous closure of ventricular septal defects. This study reports our medium term follow up of perimembranous and muscular ventricular septal defects with tunnel shape aneurysm closure using the Amplatzer duct occluder. Materials and Methods: From May 2006-December 2012, we used Amplatzer duct occluder in sev...
متن کاملTranscatheter closure of persistent ductus arteriosus in infants using the Amplatzer duct occluder.
AIM To evaluate whether transcatheter closure with the Amplatzer duct occluder offers an alternative to surgical treatment in infants with a persistent ductus arteriosus. METHODS 12 patients under 1 year of age (age 1-11 months, body weight 2.6-8.7 kg) with clinical and echocardiographic findings of a significant duct were considered for transcatheter closure with the Amplatzer occluder. The ...
متن کاملTranscatheter closure of left coronary cameral fistula with Amplatzer duct occluder II.
The surgical and transcatheter coil closure of coronary arterial fistulas are described in the literature. We report our experience with the successful transcatheter closure of a coronary arterial fistula, arising from the left coronary artery and draining into the right ventricle, with the new Amplatzer duct occluder II device.