Comparison of an Accelerometer-Based Portable Navigation System, Patient-Specific Instrumentation, and Conventional Instrumentation for Femoral Alignment in Total Knee Arthroplasty
نویسندگان
چکیده
Purpose The KneeAlign2 (KA2, OrthoAlign Inc.) accelerometer-based portable navigation system and patient-specific instrumentation (PSI; Signature, ZimmerBiomet) are widely used for ideal femoral component alignment in total knee arthroplasty (TKA). However, there has been no comparative study of the KA2 system, PSI, and conventional intramedullary instrumentation (CON). The purpose of this study was to compare the accuracy in achieving proper femoral component alignment and clinical features by using the KA2 navigation system, PSI, and CON. Materials and Methods We retrospectively compared the accuracy of femoral component alignment of 34 TKAs performed with the KA2 system for implantation of the femoral component, 32 TKAs with PSI, and 33 TKAs with CON. Results In the coronal plane, use of the KA2 system was more likely to result in optimal femoral component alignment than the CON and PSI (p<0.01). In the sagittal plane, use of the KA2 system was more likely to result in optimal component alignment than PSI, but the difference between the KA2 and CON was insignificant. Conclusions The portable accelerometer-based KA2 navigation system enabled ideal femoral implantation in the coronal and sagittal planes, as compared to the PSI or CON.
منابع مشابه
Comparison of Clinical, Functional, and Radiological Outcomes of Total Knee Arthroplasty Using Conventional and Patient-Specific Instrumentation
Background: Recently, patient-specific instrumentation (PSI) systems have been developed in order to increase theaccuracy of component positioning during total knee arthroplasty (TKA); however, the findings of previous studies arecontroversial in this regard. In the current randomized clinical study, the outcomes of computer tomography (CT)-basedPS (patient specific)-guided TK...
متن کاملUsing Accelerometer-Based Portable Navigation to Perform Accurate Total Knee Arthroplasty Bone Resection in Asian Patients.
Implant alignment is an important factor affecting clinical results associated with total knee arthroplasty (TKA). No report exists showing the utility of an accelerometer-based portable navigation system among patients with marked femoral bowing. The aim of this study was to evaluate the accuracy of a portable navigation system for implant alignment in Asian patients with marked femoral bowing...
متن کاملPatient-specific instrumentation does not improve radiographic alignment or clinical outcomes after total knee arthroplasty
Background and purpose - Patient-specific instrumentation (PSI) for total knee arthroplasty (TKA) has been introduced to improve alignment and reduce outliers, increase efficiency, and reduce operation time. In order to improve our understanding of the outcomes of patient-specific instrumentation, we conducted a meta-analysis. Patients and methods - We identified randomized and quasi-randomized...
متن کاملComputer-assisted navigation of total knee arthroplasty for osteoarthritis in a patient with severe posttraumatic femoral deformity.
In the setting of extraarticular deformities of the knee, total knee arthroplasty (TKA) is difficult, as anatomical abnormalities obstruct identification of alignment landmarks and may preclude the use of traditional instrumentation. The long-term clinical value of computer assistance for TKA is a point of ongoing controversy. Few reports describe the use of computer-assisted orthopedic surgery...
متن کاملAccuracy of patient-specific instrumentation compared with conventional instrumentation in total knee arthroplasty.
Patient-specific instrumentation (PSI) has recently been introduced to improve the alignment following total knee arthroplasty (TKA). However, controversy remains between PSI and conventional instrumentation. The aim of this study is to compare the accuracy of PSI with conventional instruments for total knee arthroplasty (TKA). A systematic literature search was performed in databases including...