A compliant, banded outflow cannula for decreased afterload sensitivity of rotary right ventricular assist devices.
نویسندگان
چکیده
Biventricular support with dual rotary ventricular assist devices (VADs) has been implemented clinically with restriction of the right VAD (RVAD) outflow cannula to artificially increase afterload and, therefore, operate within recommended design speed ranges. However, the low preload and high afterload sensitivity of these devices increase the susceptibility of suction events. Active control systems are prone to sensor drift or inaccurate inferred (sensor-less) data, therefore an alternative solution may be of benefit. This study presents the in vitro evaluation of a compliant outflow cannula designed to passively decrease the afterload sensitivity of rotary RVADs and minimize left-sided suction events. A one-way fluid-structure interaction model was initially used to produce a design with suitable flow dynamics and radial deformation. The resultant geometry was cast with different initial cross-sectional restrictions and concentrations of a softening diluent before evaluation in a mock circulation loop. Pulmonary vascular resistance (PVR) was increased from 50 dyne s/cm(5) until left-sided suction events occurred with each compliant cannula and a rigid, 4.5 mm diameter outflow cannula for comparison. Early suction events (PVR ∼ 300 dyne s/cm(5) ) were observed with the rigid outflow cannula. Addition of the compliant section with an initial 3 mm diameter restriction and 10% diluent expanded the outflow restriction as PVR increased, thus increasing RVAD flow rate and preventing left-sided suction events at PVR levels beyond 1000 dyne s/cm(5) . Therefore, the compliant, restricted outflow cannula provided a passive control system to assist in the prevention of suction events with rotary biventricular support while maintaining pump speeds within normal ranges of operation.
منابع مشابه
Passive control of a biventricular assist device with compliant inflow cannulae.
Rotary ventricular assist device (VAD) support of the cardiovascular system is susceptible to suction events due to the limited preload sensitivity of these devices. This may be of particular concern with rotary biventricular support (BiVAD) where the native, flow balancing Starling response is diminished in both ventricles. The reliability of sensor and sensorless-based control systems which a...
متن کاملA simplified cardiopulmonary bypass technique for animal experiments on implantable ventricular assist devices.
We have developed and report on a simplified cardiopulmonary bypass technique for experiments on implantable ventricular assist devices in calves. We used an electromechanical implantable ventricular assist device with a double cylindrical cam in three calves. Cannulas for the ventricular assist system were designed to be inserted between the left atrium and the descending aorta. We used the ou...
متن کاملBiventricular circulatory support with two miniaturized implantable assist devices.
BACKGROUND Up to 30% of patients with end-stage heart failure experience biventricular failure that requires biventricular mechanical support. For these patients, only bulky extracorporeal or implantable displacement pumps or the total artificial heart have been available to date, which enables only limited quality of life for the patients. It was our goal to evaluate a method that would allow ...
متن کاملCardiac Transplantation and Surgery for Heart Failure Biventricular Circulatory Support With Two Miniaturized Implantable Assist Devices
Background—Up to 30% of patients with end-stage heart failure experience biventricular failure that requires biventricular mechanical support. For these patients, only bulky extracorporeal or implantable displacement pumps or the total artificial heart have been available to date, which enables only limited quality of life for the patients. It was our goal to evaluate a method that would allow ...
متن کاملEvaluation of bend relief disconnection in patients supported by a HeartMate II left ventricular assist device.
L eft ventricular assist devices (LVADs) are an established treatment for patients with end-stage heart failure as either a bridge to cardiac transplantation 1 or as lifelong support, also known as destination therapy.) is a continuous flow device that was approved by the US Food and Drug Administration in 2008, after a pivotal trial in 133 patients awaiting transplantation. 3 More patients hav...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Artificial organs
دوره 39 2 شماره
صفحات -