Load-bearing capacity of screw-retained CAD/CAM-produced titanium implant frameworks (I-Bridge®2) before and after cyclic mechanical loading
نویسندگان
چکیده
UNLABELLED Implant-supported screw-retained fixed dental prostheses (FDPs) produced by CAD/ CAM have been introduced in recent years for the rehabilitation of partial or total endentulous jaws. However, there is a lack of data about the long-term mechanical characteristics. OBJECTIVE The aim of this study was to investigate the failure mode and the influence of extended cyclic mechanical loading on the load-bearing capacity of these frameworks. MATERIAL AND METHODS Ten five-unit FDP frameworks simulating a free-end situation in the mandibular jaw were manufactured according to the I-Bridge®2-concept (I-Bridge®2, Biomain AB, Helsingborg, Sweden) and each was screw-retained on three differently angulated Astra Tech implants (30º buccal angulation/0º angulation/30º lingual angulation). One half of the specimens was tested for static load-bearing capacity without any further treatment (control), whereas the other half underwent five million cycles of mechanical loading with 100 N as the upper load limit (test). All specimens were loaded until failure in a universal testing machine with an occlusal force applied at the pontics. Load-displacement curves were recorded and the failure mode was macro- and microscopically analyzed. The statistical analysis was performed using a t-test (p=0.05). RESULTS All the specimens survived cyclic mechanical loading and no obvious failure could be observed. Due to the cyclic mechanical loading, the load-bearing capacity decreased from 8,496 N±196 N (control) to 7,592 N±901 N (test). The cyclic mechanical loading did not significantly influence the load-bearing capacity (p=0.060). The failure mode was almost identical in all specimens: large deformations of the framework at the implant connection area were obvious. CONCLUSION The load-bearing capacity of the I-Bridge®2 frameworks is much higher than the clinically relevant occlusal forces, even with considerably angulated implants. However, the performance under functional loading in vivo depends on additional aspects. Further studies are needed to address these aspects.
منابع مشابه
Concepts for designing and fabricating metal implant frameworks for hybrid implant prostheses.
Edentulous patients have reported difficulties in managing complete dentures; they have also reported functional concerns and higher expectations regarding complete dentures than the dentists who have treated them. Some of the objectives of definitive fixed implant prosthodontic care include predictable, long-term prostheses, improved function, and maintenance of alveolar bone. One of the keys ...
متن کاملInfluence of the implant abutment types and the dynamic loading on initial screw loosening
PURPOSE This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS Three groups of abutments were produced using different types of fabrication methods; stock abutment, ...
متن کاملEffect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis
PURPOSE The purpose of this study was to examine the abutment screw stability of screw- and cement-retained implant-supported dental prosthesis (SCP) after simulated cement washout as well as the stability of SCP cements after complete loosening of abutment screws. MATERIALS AND METHODS Thirty-six titanium CAD/CAM-made implant prostheses were fabricated on two implants placed in the resin mod...
متن کاملA comparative study of gold UCLA-type and CAD/CAM titanium implant abutments
PURPOSE The aim of this study was to evaluate the interface accuracy of computer-assisted designed and manufactured (CAD/CAM) titanium abutments and implant fixture compared to gold-cast UCLA abutments. MATERIALS AND METHODS An external connection implant system (Mark III, n=10) and an internal connection implant system (Replace Select, n=10) were used, 5 of each group were connected to mille...
متن کاملLoad-bearing capacity of all-ceramic three-unit fixed partial dentures with different computer-aided design (CAD)/computer-aided manufacturing (CAM) fabricated framework materials.
The purpose of this in vitro study was to compare the load-bearing capacity of posterior three-unit fixed dental prostheses (FDP) produced with three different all-ceramic framework materials: glass-infiltrated alumina (ICA), glass-infiltrated alumina strengthened with zirconia (ICZ), and yttria-stabilized polycrystalline zirconia (YZ). Additionally, the influence on aging of mechanical cyclic ...
متن کامل