Holding power of variable pitch screws in osteoporotic, osteopenic and normal bone: are all screws created equal?
نویسندگان
چکیده
INTRODUCTION Biomechanical properties of four different commercially available small fragment cannulated screws (Twin fix (Stryker, Freiburg, Germany), Herbert, (Zimmer, Warsaw, USA), Omnitech (Unimedical, Torino, Italy), Barouk (Depuy, Warsaw, USA)), with variable pitch, used for fracture fixation were compared. MATERIALS AND METHODS Polyurethane foam blocks of three different densities with mechanical properties similar to osteoporotic, osteopenic and normal bones were used to conduct the tests. Each screw was tested for pushout and pullout holding power after a primary insertion and for pullout after a repeated insertion into the respective foam blocks. RESULTS The mean pullout and pushout strengths of all screws correlated to the foam density, and were significantly (p<0.001 and <0.001, respectively) better in foam with higher density. The mean pullout strength of each screw was consistently lower after reinsertion into the osteoporotic, osteopenic and normal bone densities by 4-30%, when compared to the index insertion (Fig. 4b). Yet, this difference was not found to be statistically significant (p=0.23). The Barouk screw performed significantly (p<0.0001) better than the other screws in all three different densities of foam for both for pushout and pullout after index insertion as well as for pullout tests after reinsertion. CONCLUSION The holding power of screws is directly correlated to bone density, thread design and number of threads engaging the bone. Reinsertion through the same hole could reduce the ultimate pullout strength. The surgeon should consider the advantages and disadvantages of each implant, depending on the clinical situation and choose accordingly.
منابع مشابه
Thread Pitch Variant in Orthodontic Mini-screws: A 3-D Finite Element Analysis
Orthodontic miniscrews are widely used as temporary anchorage devices to facilitate orthodontic movements. Miniscrew loosening is a common problem, which usually occurs during the first two weeks of treatment. Macrodesign can affect the stability of a miniscrew by changing its diameter, length, thread pitch, thread shape, tapering angle and so on. In this study, a 3-D finite element analysis wa...
متن کاملFracture fixation with two locking screws versus three non-locking screws
OBJECTIVES We aimed to further evaluate the biomechanical characteristics of two locking screws versus three standard bicortical screws in synthetic models of normal and osteoporotic bone. METHODS Synthetic tubular bone models representing normal bone density and osteoporotic bone density were used. Artificial fracture gaps of 1 cm were created in each specimen before fixation with one of two...
متن کاملمقایسه شاخصهای پایداری ایستا در زنان سالم، استئوپنیک و استئوپروتیک با استفاده از نوسان مرکز فشار(Center Of Pressure)
Objective: Change of control strategy in osteoporotic people and their focus on the hip strategy, also the occurred postural changes, result in the displacement of center of pressure (COP). This study was carried out to compare the static stability parameters in healthy, osteopenic and osteoporotic women. Methods: Seventy-one volunteer postmenopausal women aged 48-65 years were divided into th...
متن کاملA comparison of screw insertion torque and pullout strength.
PURPOSE Pullout strength of screws is a parameter used to evaluate plate screw fixation strength. However, screw fixation strength may be more closely related to its ability to generate sufficient insertion because stable nonlocked plate-screw fracture fixation requires sufficient compression between plate and bone such that no motion occurs between the plate and bone under physiological loads....
متن کاملEvaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method
longevity of osseointegrated implants are intensely influenced by biomechanical factors. Control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. In this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum Von Mises stress in implant-abutment complex and ja...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Injury
دوره 41 2 شماره
صفحات -
تاریخ انتشار 2010