Characterization of an x-ray phase contrast imaging system based on the miniature synchrotron MIRRORCLE-6X.
نویسندگان
چکیده
PURPOSE The implementation of in-line x-ray phase contrast imaging (PCI) for soft-tissue patient imaging is hampered by the lack of a bright and spatially coherent x-ray source that fits into the hospital environment. This article provides a quantitative characterization of the phase-contrast enhancement of a PCI system based on the miniature synchrotron technology MIRRORCLE-6X. METHODS The phase-contrast effect was measured using an edge response of a plexiglass plate as a function of the incident angle of radiation. We have developed a comprehensive x-ray propagation model based on the system's components, properties, and geometry in order to interpret the measurement data. Monte-Carlo simulations are used to estimate the system's spectral properties and resolution. RESULTS The measured ratio of the detected phase-contrast to the absorption contrast is currently in the range 100% to 200%. Experiments show that with the current implementation of the MIRRORCLE-6X, a target smaller than 30-40 μm does not lead to a larger phase-contrast. The reason for this is that the fraction of x-rays produced by the material (carbon filament and glue) that is used for mounting the target in the electron beam is more than 25% of the total amount of x-rays produced. This increases the apparent source size. The measured phase-contrast is at maximum two times larger than the absorption contrast with the current set-up. CONCLUSIONS Calculations based on our model of the present imaging system predict that the phase-contrast can be up to an order of magnitude larger than the absorption contrast in case the materials used for mounting the target in the electron beam do not (or hardly) produce x-rays. The methods described in this paper provide vital feedback for guiding future modifications to the design of the x-ray target of MIRRORCLE-type system and configuration of the in-line PCI systems in general.
منابع مشابه
Characterization of the Phase Contrast Imaging system based on the miniature synchrotron MIRRORCLE-6X
miniature synchrotron MIRRORCLE-6X Joop van Heekeren, a) Alexander Kostenko, a) Takayasu Hanashima, Hironari Yamada, Sjoerd Stallinga, S. Erik O erman, and Lucas van Vliet 1)Quantitative Imaging Group, Department of Applied Sciences, Delft University of Technology, Lorentzweg 1, 2628 CJ Delft, Netherlands 2)Photon Production Laboratory Ltd, Ritsumeikan University, 1-1-1 Noji-higashi, Kusatsu, S...
متن کاملHigh Resolution X-ray Imaging by Portable Synchrotron Radiation Source "mirrorcle-6x"
MIRRORCLE-6X is a portable synchrotron composed of a 6-MeV microtron injector, and a 60cm outer diameter exactly circular synchrotron ring made of a normal conducting magnet. The injection is performed at 400 Hz repetitions by 100 mA injector peaks current, which lead to 3A accumulated peak current. X-rays are generated by a collision of the relativistic electron beam and a small target placed ...
متن کاملA Review of the Applications of Synchrotron Radiation in Archaeological Sciences
The scientific research regarding investigation, characterization and protection of the archeological specimens is manifested through a notable participation of multidisciplinary subjects and experts, scientists and archeometrists. One of the main principals which are considered by archaeometrists in the study of the precious specimens is the utilizing nondestructive methods. As an example, in ...
متن کاملSynthesis and characterization of Gd2O2 S: Tb3+ phosphor powder for X-ray imaging detectors
Gadolinium oxysulfide phosphor doped with trivalent terbium have been synthesized using urea homogenous precipitation and followed by sulfurization at 800 °C under argon atmosphere. Structural and morphological of synthesized phosphor powder were characterized by x-ray diffraction (XRD), scanning electron microscopy (SEM) and Fourier transform infrared spectrometry (FT-IR). Hexagonal structure ...
متن کاملEvaluation of the Effect of Source Geometry on the Output of Miniature X-ray Tube for Electronic Brachytherapy through Simulation
Objective: The use of miniature X-ray source in electronic brachytherapy is on the rise so there is an urgent need to acquire more knowledge on X-ray spectrum production and distribution by a dose. The aim of this research was to investigate the influence of target thickness and geometry at the source of miniature X-ray tube on tube output.Method: Five sources were simulated based on problems e...