Accuracy evaluation of the primary frequency standard NIST-7

نویسندگان

  • J. H. Shirley
  • W. D. Lee
  • R. E. Drullinger
چکیده

We develop a comprehensive approach to the evaluation of the uncertainty of a primary frequency standard. Indirect, model-dependent means are employed for the evaluation of all known biases. This approach leads to a lower combined standard uncertainty (CSU) and leaves frequency measurements under altered conditions as a powerful search technique for unknown biases. We report the results of this evaluation technique applied to NIST-7, one of the United States’ primary frequency standards. The best CSU value we have reported to the Bureau International des Poids et Mesures (BIPM) is , while a rigorous application of this evaluation process would suggest that a limiting Type B uncertainty of is possible. We discuss this difference and the implications of reporting optimistic values. We also discuss the history of reported evaluations of NIST-7 including those made before the full implementation of the techniques presented here.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Th European Frequency and Time Forum Neuchatel - 16 - 17 - 18 March 1993 Nist - 7 , the New Us Primary Frequency Standard

NIST-7, an optically pumped, cesium-beam frequency standard has replaced NBS-6 as the official US primary frequency standard. The present short-term stability of the standard, measured with respect to an active hydrogen maser, is characterized by uy(t) L 8 x i". Our first evaluation has resulted in an uncertainty of 4 x An improved servoelectronic system is being developed and this should impro...

متن کامل

NIST-F1: recent improvements and accuracy evaluations∗

In the last several years we have made many improvements to NIST-F1 (a laser-cooled Cs fountain primary frequency standard at the National Institute of Standards and Technology (NIST) in Boulder, Colorado) resulting in a reduction by over a factor of 2 in the uncertainty of the realization of the SI second. The two most recent accuracy evaluations of NIST-F1 had combined standard fractional unc...

متن کامل

Accuracy evaluation of NIST-F1

The evaluation procedure of a new laser-cooled caesium fountain primary frequency standard developed at the National Institute of Standards and Technology (NIST) is described. The new standard, NIST-F1, is described in some detail and typical operational parameters are discussed. Systematic frequency biases for which corrections are made – second-order Zeeman shift, black-body radiation shift, ...

متن کامل

First accuracy evaluation of NIST-F2

We report the first accuracy evaluation of NIST-F2, a second-generation laser-cooled caesium fountain primary standard developed at the National Institute of Standards and Technology (NIST) with a cryogenic (liquid nitrogen) microwave cavity and flight region. The 80 K atom interrogation environment reduces the uncertainty due to the blackbody radiation shift by more than a factor of 50. Also, ...

متن کامل

RECENT IMPROVEMENTS IN NIST-F1 AND RESULTING ACCURACIES OF δf/f<7× 10 †

Since our last publication describing the accuracy budget for NIST-F1 (a laser-cooled Cs fountain primary frequency standard) we have made many improvements resulting in nearly a factor of 2 reduction in the uncertainty in the realization of the SI second. Recently we reported a combined standard fractional uncertainty of 0.67×10.