Experience with a new device for clearing mucus from the endotracheal tube.
نویسندگان
چکیده
Partial or total obstruction of an endotracheal tube (ETT) by mucus can cause severe respiratory distress, hypoxemia, or death. Signs of an obstructed ETT include increased ventilation pressure, changes in the ventilator graphics, S(pO(2)) decrease, and cardiovascular changes. We present 3 patients whose ETTs were partially obstructed by mucus. In each case the patient displayed adverse effects from the obstruction, but once the obstruction was removed they showed dramatic improvement. In each case we used a new device (Rescue Cath, Omneotech, Tavernier, Florida) designed to remove mucus from the ETT lumen. The 3 cases demonstrate that the device is effective and capable of relieving the adverse effects of ETT mucus obstruction.
منابع مشابه
Comparison of tracheal tube cuff inflation pressure among anesthesia team with different gradual level
Background & Objectives: Tracheal intubation is frequently performed in operating room. Over inflation of the endotracheal tube cuff is a risk factor for tracheal ischemia and subsequent complications. On the other hand inadequate cuff pressure increases the risk of aspiration of gastric contents. In this study, the endotracheal tube cuff pressure filled by the anesthesia team has been compared...
متن کاملNovel system for complete removal of secretions within the endotracheal tube: the Mucus Shaver.
THE method normally used to clean endotracheal tubes (ETTs) in intubated patients is via a small, flexible, plastic suction catheter. The small-bore tube is inserted into the ETT and manipulated to remove mucus lodged within the ETT and this process is repeated as frequently as indicated. This may not remove all secretions. With residual secretions retained on the surface, the ETT provides a fe...
متن کاملNew Metal Shield for the Endotracheal Tube during Laser Endolaryngeal Surgery
Introduction: To present and assess a newly designed autoclavable, reusable tube shield for an endotracheal tube (ETT) during laser laryngeal surgery. Materials and Methods: This study included patients scheduled for endolaryngeal laser surgery. A carbonated stainless-steel hemisphere shield (~1.5×0.6 mm) connected to a silk thread passing through a hole in its middle near upper edge was ...
متن کاملEffect of Stabilizing Endotracheal Tube on Quality of Airway Management in Patients Hospitalized in ICU
Background: The unplanned departure of endotracheal tube is one of the main complications for endotracheal intubation. In addition to endotracheal extubation, its movement can also lead to damage to the oral cavity. Stabilizing endotracheal tube is the most important factor that can prevent unplanned departure of endotracheal tube as well as its movement. The current study was done with the aim...
متن کاملEfficacy Study of a Novel Device to Clean the Endotracheal Tube
The Endotracheal Tube (ETT) is a life-saving device for respiratory support and airway protection. However, its sustained presence in the trachea disrupts the physiological mechanisms that maintain mucus homeostasis. The process of mucus buildup and luminal narrowing leads to an increased ventilatory effort and might delay liberation from mechanical ventilation. In addition, potential lung path...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 56 4 شماره
صفحات -
تاریخ انتشار 2011