Robust and Automatic Calibration Method for 3D Freehand Ultrasound
نویسندگان
چکیده
This paper describes a new robust and fully automatic method for calibration of three-dimensional (3D) freehand ultrasound. 3D Freehand ultrasound consists in mounting a position sensor on a standard probe. The echographic B-scans can be localized in 3D, and can be compounded into a volume. However, especially for quantitative use, this process dramatically requires a calibration procedure that determines its accuracy and usefullness. Calibration aims at determining the transformation (translations, rotations, scaling) between the coordinates system of the echographic images and the coordinate system of the localization system. To calibrate, we acquire images of a phantom whose 3D geometrical properties are known. We propose a robust and fully automatic calibration method based on the Hough transform and robust estimators. Experiments have been done with synthetic and real sequences, and this calibration method is shown to be easy to perform, accurate, automatic and fast enough for clinical use.
منابع مشابه
A fully automatic calibration procedure for freehand 3D ultrasound
This paper describes a novel method for calibration of freehand three-dimensional (3D) ultrasound. A position sensor is mounted on a conventional ultrasound probe, thus the set of B-scans can be localized in 3D, and can be compounded into a volume. The calibration process aims at determining the transformation (translations, rotations, scaling) between the coordinates system of images and the c...
متن کاملCalibration of Swept-Volume 3D Ultrasound
We present two different methods for calibrating 3D swept-volume ultrasound as a prerequisite for an ultrasound based intraoperative navigation system. Different solutions of the calibration problem for 2D ultrasound exist, but no approach has been published so far that takes advantage of 3D swept-volume ultrasound. In contrast to 3D freehand ultrasound no compounding has to be performed. Inste...
متن کاملImage-Based Method for In-Vivo Freehand Ultrasound Calibration
For freehand ultrasound systems, a calibration method is necessary to locate the position and orientation of a 2D B-mode ultrasound image plane with respect to a position sensor attached to the transducer. In addition, the acquisition time discrepancy between the position measurements and the image frames has the be computed. We developed a new method that adresses both of these problems, based...
متن کاملUltrasound Guided Surgery: Image Processing and Navigation
The process of 3D freehand ultrasound acquisition consists of two steps: scanning and reconstruction. Prior to scanning, a position sensor is attached to the probe for tagging each image with its position and orientation in space. A crucial step prior to reconstruction and visualization is calibration of the ultrasound probe with the attached position sensor, i.e. determining the position and o...
متن کاملUltrasound Calibration Toolkit with a High-Adjustability Feature Based on User Requirements
3D freehand Ultrasound reconstruction is one of many choices to assemble 3D ultrasound volume from 2D ultrasound images with benefit in low-cost operations, conveniency and ease of usage which can be applied to intraoperative imaging. Ultrasound calibration is the procedure used to find the position and orientation of each acquired 2D ultrasound image in 3D space, hence is needed to be performe...