Evaluation of the Nova Biomedical Statstrip glucose meter
نویسنده
چکیده
Aims: The Nova StatStrip blood glucose monitoring system was evaluated for precision, accuracy, and interferences from haematocrit and maltose. It was also compared against three other meters and two reference methods. Methods: Heparinised whole blood samples were analysed on the meters. These results were compared with whole blood samples analysed on the Radiometer ABL835 for the interference studies. Plasma samples, obtained from these whole blood samples, were measured on the Abbott CI8200 for accuracy studies. Results: There were signifi cant differences in the degree to which the meters correlated with the reference method. With the exception of the StatStrip, all meters were affected by variable haematocrit. Of the two glucose meters tested, the StatStrip did not show any maltose interference. Conclusions: The StatStrip glucose meter did not show clinically signifi cant interference from maltose or varying haematocrit levels. In addition, the StatStrip demonstrated the best correlation with the reference glucose method.
منابع مشابه
Performance of the Statstrip Glucose Meter in Inpatient Management of Diabetes Mellitus
The use of hospital point-of-care glucose meters in inpatient care is widely established despite unsolved issues regarding accuracy. The aim of this study was to validate the performance of the StatStrip Glucose Hospital Meter (Nova Biomedical, Waltham, USA) in performing glycemic profile in hospitalized diabetic patients undergoing intensified insulin therapy. We investigated total imprecision...
متن کاملValidation of Point-of-Care Glucose Testing for Diagnosis of Type 2 Diabetes
Point-of-care (POC) glucose technology is currently considered to be insufficiently accurate for the diagnosis of diabetes. The objective of this study was to investigate the diagnostic accuracy of an innovative, interference-resistant POC glucose meter (StatStrip glucose hospital meter, Nova Biomedical, USA) in subjects with a previous history of dysglycaemia, undergoing a 75 g diagnostic oral...
متن کاملAccuracy and reliability of the Nova StatStrip® glucose meter for real-time blood glucose determinations during glucose clamp studies.
AIMS/HYPOTHESIS The Andres clamp technique, which requires accurate and timely determination of glucose, utilizes the Beckman or Yellow Springs Instruments (YSI) glucose analyzers. Both instruments require maintenance, a dedicated operator, preparation of a plasma sample, and a duplicate measurement that takes ≥2 minutes. The Nova StatStrip glucose meter was evaluated for accuracy, reliability,...
متن کامل[Validation of a glucose meter at an intensive care unit].
INTRODUCTION Hyperglycemic patients admitted to Intensive care units (ICUs) have higher morbidity and mortality than normoglycemic patients. Blood glucose levels of ICU patients are usually measured with a glucose meter. The aim of this study was to evaluate a glucose meter (StatStrip, Nova Biomedical) to assess its agreement with the standard laboratory method for testing glucose. MATERIAL A...
متن کاملPoint - f - c re glucose mo it ring on the eonatal unitjpc
Aim: This study aims to compare the accuracy and precision of the Nova StatStrip glucometer against the Radiometer ABL gas analyser. Based on the results, to establish if the Nova StatStrip glucometer could be adopted as a reliable alternative for near-patient glucose monitoring on the neonatal unit. Methods: Seven hundred twenty-eight paired samples were collected prospectively from babies on ...