Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner
نویسندگان
چکیده
PURPOSE This retrospective study was performed to evaluate the clinical results and measure polyethylene liner wear in total hip arthroplasty (THA) with highly cross-linked polyethylene. MATERIALS AND METHODS Except for patients who had died or were unable to have follow-up at least 2 years, 60 of 78 hips that underwent THA were included this study. The mean age was 64.5 years (range, 25-81 years) and the mean body mass index (BMI) was 23.0 kg/m(2) (18.1-32.3 kg/m(2)). Diagnosis at the time of the operation was osteonecrois of the femoral head in 28 hips, primary osteoarthritis in 14, hip fracture in 13, and other diseases in 5. The mean follow-up period was 3.8 years (2.1-7.1 years). Harris hip score (HHS) was reviewed before THA and at the last follow-up. On the anteroposterior pelvic radiographs, acetabular cup inclination and ante-version were also measured. The annual linear wear rate was measured using Livermore's method on the radiographs. RESULTS The mean HHS was 60.1 (28-94) before operation and 90.4 (47-100) at the last follow-up. In the immediate post-operation, the average inclination and anteversion angles of the acetabular cups were 46.3° (standard deviation, ±6.7°) and, 21.4°(±10.1°) respectively. The mean of the annual linear polyethylene wear was 0.079 mm/year (0.001-0.291 mm/year). Age, gender and BMI were not statistically related to linear polyethylene wear but the period of follow-up and the acetabular cup's inclination showed significant negative and positive correlation respectively. CONCLUSION The wear rate of a highly cross-linked polyethylene was shown to correlate negatively with duration of follow-up. However, our study was based on a short-term follow-up, so a long-term follow-up study is necessary in the future.
منابع مشابه
A Liner Breakage in Total Hip Arthroplasty after Using 1st Generation Highly Cross Linked Polyethylene Mated against 36-mm Metal Head: A Case Report
It has been known the highly cross linked polyethylene (HXLPE) has an advantage of improved wear rate. However, the alteration in mechanical properties such as decreased tensile yield and fatigue strength make concerns about fragility of HXLPE. We experienced a case of HXLPE breakage. But, this case of liner breakage happened although patient belonged to normal BMI and proper acetabular cup pos...
متن کاملIncarcerated fracture fragments of Longevity polyethylene liners after total hip arthroplasty
Highly cross-linked polyethylene liners are widely used in total hip arthroplasty because they experience lower wear rates than conventional polyethylene liners. However, the cross-linking process does decrease the resistance of polyethylene to fatigue failure and fracture. This report describes 2 cases of highly cross-linked polyethylene liner fracture occurring in association with hip disloca...
متن کاملCatastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty
INTRODUCTION Catastrophic polyethylene failure is a rare complication of ceramic-on-polyethylene total hip arthroplasty due to the favorable tribological characteristics of ceramic. Failure of the polyethylene liner can be disastrous, increasing periprosthetic osteolysis, metallosis, and risk of dislocation. Complications associated with ceramic-on-polyethylene articulations have been studied e...
متن کاملOne-component revision of failed hip resurfacing from adverse reaction to metal wear debris.
This study assessed the results of 90 one-component revisions for failed hip resurfacing due to adverse reaction to metal wear debris (76 acetabular, 14 femoral). Patients with a femoral head size 40-45 mm (n=33) received a two-piece titanium meshed shell with a cross-linked polyethylene liner and patients with femoral head size 46-54 mm (n=43) received metal-on-metal components. Patients with ...
متن کاملHighly Cross-linked Polyethylene Liner Dissociation from a Cement-less Modular Acetabular Shell: Two Case Reports
Liner dissociation of polyethylene from a cementless acetabular socket following total hip arthroplasty (THA) is a rare complication. Cross-linked polyethylene liner dissociation from AMS-HA shell (KYOCERA Med, Osaka, Japan) occurred in 2 out of the 4153 (0.04%) cases approximately 10 years after undergoing surgery at our institute. First case was an 80-year-old female who underwent right THA a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 27 شماره
صفحات -